2-8

Smart average vs. normal average

Normal average

 smoothes the waveform of repetitive

signals.  It gives you the averaged waveform over
successive acquisitions.  A single waveform that deviates
strongly will affect the smoothed waveform.  This happens
for example when looking at an interlaced video system.
The fly-back line can distort the waveform, and you will not
see it flash by as a separate waveform (See Figure 2-5).

Another characteristic of normal average is that when the
signal really changes, for example if you are probing
around, it takes some time before the new signal becomes
stable.  The displayed waveform slowly changes from one
waveform to another waveform.

Smart average

 gives you the primary waveform, but

displays an incidental trace of different waveshape directly
and  without affecting the primary waveform .
You can now for example look at the averaged waveform
on an interlaced video system, and still see the fly-back
line flash by (See Figure 2-6).

When the signal changes are of more permanent nature,
the averaging algorithm automatically restarts and
displays the new waveform instantly, without displaying
the distorted intermediate waveshapes.

Figure 2-5. Normal Average

Figure 2-6. Smart Average

Summary of Contents for ScopeMeter 190B Series

Page 1: ...4822 872 00741 Nov 2002 Rev 1 3 04 2002 2003 Fluke Corporation All rights reserved All product names are trademarks of their respective companies Fluke 190B C Medical Functions Users Manual Supplement...

Page 2: ...cal Version 1 2 MA 190 Accessory Kit 1 2 2 Using the Medical Functions 2 1 mAs Measurements 2 1 mVs measurements 2 3 mWs measurements 2 3 Extended Offset 2 4 Oscilloscopes and dynamic range 2 5 Trigge...

Page 3: ...ltage 42 Vpk use only and do not comply with the voltage rating on most other ScopeMeter accessories Carefully read the warning below which is an explicit addition to the general accessory and ScopeMe...

Page 4: ...NC feed through terminator for video and HF measurements Max 5 V See Fig 1 1 TA 190 50R 50 Ohm 10 1 attenuator feed through terminator for additional dynamic range and offset See Fig 1 1 CS 190 1R 1 O...

Page 5: ...frozen waveforms HOLD recorded waveforms and on saved waveforms To perform mAs measurements on input A do the following 1 Connect the current shunt CS190 1R to input A using the BNC cable see Figure...

Page 6: ...ding mA or A A etc the difference between the current amplitude measured on the left cursor and measured on the right cursor Top right reading mAs or As As etc the integrated current over the time bet...

Page 7: ...els B and C versions are different 8 Press to highlight Observe that two vertical cursors are displayed 9 Press to highlight mAs 10 Highlight the left or right cursor 11 Move the highlighted cursor to...

Page 8: ...he signal To choose the extended offset range for input A do the following 1 Display the INPUT A key labels 2 Open the INPUT A menu 3 2x Jump to Offset range 4 Select Extended and accept the extended...

Page 9: ...ust take care when using high time base speeds when looking at signals that have large very steep edges They can be distorted when they jump from outside the dynamic range directly to the visible area...

Page 10: ...fied Pulse Trigger mode Tip for video signal measurements To gain full profit of the Extended Offset for video signal measurements use the 50 10 1 attenuator feed through terminator select Probe on A...

Page 11: ...e when you probe around it takes quite some time before the new waveshape is stable With smart averaging you can quickly probe around and incidental waveform changes like a line flyback in video show...

Page 12: ...changes for example if you are probing around it takes some time before the new signal becomes stable The displayed waveform slowly changes from one waveform to another waveform Smart average gives y...

Page 13: ...ect positive signal polarity for video signals with negative going sync pulses 6 Select Non interlaced to open the Scan rate menu 7 Select the required Scan rate and confirm the selection Trigger prob...

Reviews: