RUSKA 7252
Users Manual
6-2
3.
Press
Electrical
[
F1
]. The electronics self test will run and display the results.
The electronics test runs eight sets of tests on various parts of the electronic modules.
Table 6-1 describes these tests and the possible actions needed if a test fails. The
pneumatics test runs nine tests on the pneumatics module.
Table 6-1. Electronics Self Test
Test
Description
Action on Failure
Processor
Tests the processor chip
Replace processor board.
Clock
Tests the real time clock
Replace processor board.
Timer
Tests the 10 ms interval timer
Replace processor board.
EEPROM
Tests the nonvolatile memory
Replace processor board.
Oven
Tests operation of the sensor oven
Replace sensor assy.
Allow oven to warm up.
Removing the Calibrator’s Cover
The Calibrator should be kept clean and completely assembled at all times. Operating the
Calibrator without its cover affects the Calibrator’s thermal gradients and therefore may
reduce precision. If it becomes necessary to remove the Calibrator’s cover, follow the
instructions below.
WX
Warning
The Calibrator should only be opened by qualified
electrical/mechanical service technicians. Lethal voltages are
present and exposed in the power supply and display.
1.
Turn off the Calibrator and disconnect the power cord from the Calibrator.
2.
Locate and unscrew the four screws that secure the cover to the back panel.
3.
Place your hands near the middle of the cover and slide the cover towards the
Calibrator’s back panel.
4.
Lift up on the cover. With the cover removed, use typical electronic cleaning tools to
remove any accumulated dust from inside the instrument.
5.
Replace the cover before resuming operation.
Moisture Filter
The Bourdon tube sensor is hydroscopic. An external desiccant filter prevents
introduction of moisture and is strongly recommended for high humidity areas. The filter
should be replaced annually. The moisture filter is used with the reference port only
(absolute only instruments do not need a moisture filter).
Particle Filters
Refer to Appendix A for air supply quality requirements. During normal operation, the
Calibrator transfers gas both into and out of the device under test (DUT). When necessary,
the user is expected to use and maintain an in-line disposable particle filter to protect the
Calibrator’s pneumatics from any contamination that may exist in the DUT.
Summary of Contents for RUSKA 7252
Page 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Page 4: ......
Page 10: ...RUSKA 7252 Users Manual vi ...
Page 12: ...RUSKA 7252 Users Manual viii ...
Page 20: ...RUSKA 7252 Users Manual 1 6 ...
Page 40: ...RUSKA 7252 Users Manual 3 6 ...
Page 72: ...RUSKA 7252 Users Manual 4 32 ...
Page 114: ...RUSKA 7252 Users Manual 6 22 ...
Page 118: ...RUSKA 7252 Users Manual 7 4 ...