Remote Commands
Remote Commands
9
9-5
R - Read ADC Reference Values
Command: <start><addr><Size>R<chk>
Reply:
<start><addr><Size>r<temp><rh><press><chk>
Command <Size>: 1
Reply <Size>: 7
All values are 16-bit signed integers using scaled (*100) units.
<temp> is degrees C
<rh> is %
<press> is kPa.
S - Read Remote Pressure with Displayed Precision
Command: <start><addr><Size>S<chk>
Reply:
<start><addr><Size>s<status><press><decimals><chk>
Command <Size>: 1
Reply <Size>: 9
<status> is three unsigned 8-bit values. (See Command A)
<press> is 32-bit unsigned integer using scaled (*10^decimals) units. Units is kPa.
<decimals> is 8-bit signed integer
V - Get Version Info
Command: <start><addr><Size>V<chk>
Reply: <start><addr><Size>v<versionmajor><versionminor><HWVer>
<SubModel><model><chk>
Command <Size>: 1
Reply <Size>: 7
<versionmajor> is an unsigned 8-bit unsigned value
<versionminor> is an unsigned 8-bit unsigned value
<HWVer> is an unsigned 8-bit unsigned value
<SubModel> is an unsigned 8-bit unsigned value
<model> is a 2-byte flag used to differentiate between software with different command sets.
W - Write EEPROM
Command: <start><addr><Size>W<addr1><data><chk>
Reply:
<start><addr><Size>w<data><chk>
Command <Size>: 11
Reply <Size>: 9
<addr1> is a 16-bit address, least significant byte first.
<data> is 8 bytes read from EEPROM.
To write the eeprom, set the high byte of <addr> to 00H.
To write low memory, set the high byte of <addr> to F0H.
To write extended memory, set the high byte of <addr> to F1H.
Address should be an 8-byte boundary.
Z - Zero
Command: <start><addr><Size>Z<channel><value><chk>
Reply:
<start><addr><Size>z<status><counts><chk>
Command <Size>: 4
Reply <Size>: 6
<channel> is an 8-bit value.
<channel> selects channel to be calibrated. P for pressure, H for humidity, T for temperature.
<value> is a 16-bit signed integer using scaled (*100) units matching select channel.
<status> is three unsigned 8-bit values. (See Command A)
<counts> is an unsigned 16-bit value.
Summary of Contents for RUSKA 2456-LEM
Page 6: ...RUSKA 2456 LEM Users Manual iv...
Page 8: ...RUSKA 2456 LEM Users Manual vi...
Page 10: ...RUSKA 2456 LEM Users Manual viii...
Page 24: ...RUSKA 2456 LEM Users Manual 3 2...
Page 32: ...RUSKA 2456 LEM Users Manual 6 2...