RUSKA 2456-LEM
Users Manual
2-6
Viewing Real Time Data
gky11.bmp
Figure 2-8. Main Screen
The Memo line allows a user to enter a descriptive to identify each LEM. This
description is associated with the Serial Number and is stored in the *.ini file on the host
computer.
The green indicator at the end of each line identifies if the LEM data is being logged.
See Chapter 2, Data Logging
*** If the Data Line turns yellow then the data being displayed is out of the following
ranges: 17
o
C < temperature <29
o
C
0% < relative humidity < 100%
10 psi < pressure < 16.7 psi
*** If the Data Line turns red then communication has been lost with the LEM.
Logging Data
The Data Logging function captures Temperature, Pressure, Relative Humidity and Air
Density (Chapter 7.1 - NBS Pub. 700-1)
Setup
From the Menu select Log (Figure 2-9).