Operation
Remote Operation via the RS-232 Function
3
3-25
Table 3-2. Available Duplex Mode Commands (cont.)
Command Description
Parameters
Required
Return Value
(Response)
RDCURRENT Prompts
an
Analyzer
current measurement in
milliamperes. Not
available in RARF
mode.
None AAAA
(mA)
Where AAAA is
current in
milliamperes.
RDMETER
Prompts the Analyzer to
measure the applied
signal of the ESU.
None
Prompts the Analyzer to
return all measurements
of the applied signal of
the ESU.
{MODE=NONE}
None I=AAAA(mA),
CF=XX.X,
KVPP=XXX,
PWR=PPP.P(w)
Where AAAA is the
Current reading in
milliamps, XX.X is the
Crest Factor, XXX is
the peak-to-peak
voltage in kVolts and
PPP.P is the power in
watts.