
Operation
Testing
the
ESU
3
3-9
To observe the ESU’s visual or audible alarm:
1.
With the 50-ohm test load attached to the ESU CQM (REM) input, it
should sense this level of resistance in the "pass zone." Depending upon
the ESU device manufacturer, this pass zone can extend from 135 to 250
ohms. Test load settings over the pass zone activate the CQM alarm.
Note
Refer to the technical support/service manual of the ESU you are
testing for the recommended check-out procedure for the ESU CQM
feature.
2.
Increase the test load from 50 ohms until the CQM alarm sounds. The
value at that point is the CQM alarm resistance.
DISPERSIVE
ACTIVE
ESU
Basic CQM Check
(Contact Quality Monitor)
RF303
X No connection
ESU CQM
BLU
YEL
Main
Test
Load
Supply
Mains
Earth
CQM
exz13.eps
Figure 3-2. Basic CQM Check
Summary of Contents for RF303 Operators
Page 8: ...RF303 Operators Manual iv...
Page 10: ...RF303 Operators Manual vi...
Page 12: ...RF303 Operators Manual 1 2...
Page 24: ...RF303 Operators Manual 1 14...
Page 26: ...RF303 Operators Manual 2 2...
Page 38: ...RF303 Operators Manual 3 2...
Page 64: ......
Page 66: ...RF303 Operators Manual A 2...
Page 68: ...RF303 Operators Manual B 2...