MicroScanner Pro™
Calibration Mode
7
If you know a cable’s NVP, change the displayed
numbers using the
keys until the appropriate
NVP is displayed. The cable length will automatically
adjust to the new NVP.
If you know a cable’s length, change the shown NVP
using the
keys until the appropriate length is
displayed. The NVP can be adjusted in 1% increments,
and the length changes accordingly.
Cables used for calibration must be at least 50 feet (15
meters) long. Cable lengths of less than 50 feet will
display
FAULT
.
Changing the Display from Meters to Feet
During Calibration you will be able to switch the
displayed length from meters to feet by simply pressing
the
MODE
key.
Press the
ON/OFF
key once the desired cable length or
NVP is displayed to terminate Calibrate mode and store
the new calibration factor. M
ICRO
S
CANNER
P
RO
will use it
for future length measurements until another
calibration is performed.