Fluke LS-1310 Information Sheet Download Page 6

LS-1310/1550
Laser Source

6

 

W

Caution

 In the following steps, do not exceed
the laser diode’s maximum operating
current of 60 mA. Doing so can damage
the laser diode. If the current reaches
60 mA before the optical power reading
reaches -9.2 dBm, clean the ST
connectors or repair the laser source.

4-4.

 

Set the laser source to 1550 nm. Verify that
the laser diode current is less than 60 mA.

 

W

Caution

In the next step, slow the adjustment
rate on R14 as the optical power reading
approaches -14 dBm. If you turn R14 too
quickly, you might overshoot the
-9.2 dBm target and exceed the laser
diode’s maximum current of 60 mA.

4-5.

 

Adjust R14 clockwise until either the power
reading is -9.2 dBm ±0.1 dB or the current
reaches 60 mA. A typical current value is
25 mA at -9.2 dBm. If the current reaches
60 mA before the power reading reaches
-9.2 dBm, the laser source requires repair.

4-6.

 

Turn off the laser source. Disconnect the
DMM. Reinstall the jumper between J2-1
and J2-2.

4-7.

 

With the laser source set to 1550 nm, verify
that the power reading is -9.2 dBm ±0.1 dB.

4-8.

 

Adjust the 1550 nm external potentiometer
(R35) counterclockwise to verify that the
laser source output is adjustable to
-11.2 dBm ±0.1 dB.

4-9.

 

Adjust R35 until the laser source output
reads -10.0 dBm ±0.1 dB.

4-10.

 

While the laser source is on, disconnect the
power supply; then install a 9 V battery.
Verify that the output is still at -10.0 dBm
± 0.1 dB.

4-11.

 

Connect an oscilloscope across J2-1 and J2-4
(Gnd). Set the laser source 

CW/MOD

 switch

to the 

MOD

 position.

4-12.

 

Verify that the power reading is -10 dBm
±0.5 dBm. Verify that the signal frequency
is 2 kHz ±100 Hz with a duty cycle of 50%
±5%.

4-13.

 

Set the laser source 

CW/MOD

 switch to 

CW

.

Turn off the laser source.

Step 5: Reassembly and Final Test

5-1.

 

Remove the singlemode cable from the laser
source. Put the dust cap onto the laser source
optical ST connector.

5-2.

 

Apply insulating varnish, such as red GLPT,
to all potentiometers adjusted.

5-3.

 

Put the board back in the case. Put the case
halves together and install the four Phillips
screws. Reinstall the lock washers and nuts to
the CW/MOD and 1550/1310 switches.
Install a 9 V battery and the battery door.

5-4.

 

Run the performance tests as described in the
earlier section "Performance Tests".

Calibration Adjustments for Serial
Numbers 

 79130000

(For earlier serial numbers, see page 3)

To ensure that the laser source performs to
specifications, calibrate it annually using the
procedure that follows.

Always perform the complete calibration
adjustment procedure.

Step 1: Preparing for Calibration

Before you start the calibration procedures
described in this sheet, do the following (refer to
Figure 2 for locations of components on the circuit
board):

1-1.

 

Turn off the laser source.

1-2.

 

Remove the battery door and the 9 V battery.

1-3.

 

Remove the four Phillips screws from the
back of the laser source. Remove the top case.

1-4.

 

Unscrew the external retaining nuts from the
laser source's power switch and 

CW/MOD

switch.

1-5.

 

Carefully remove the laser source circuit
board from the bottom case. Take care to not
strain the fiber that connects the circuit board
to the ST connector mounted in the bottom
case.

Summary of Contents for LS-1310

Page 1: ...r radiation and to prevent eye damage Never look directly into the source connector Though the laser radiation is invisible it can damage your eyes Do not modify the source Do not magnify or otherwise modify the laser output Use only approved connectors and adapters Do not use controls adjustments or procedures not documented or approved by Fluke Corporation Caution While servicing the laser sourc...

Page 2: ...ance tests and calibration adjustments Fiber optic power meter calibrated at 1310 nm and 1550 nm 0 10 V dc adjustable power supply with a minimum output of 500 mA and a current limit adjustment Fluke 187 DMM digital multimeter equivalent or better 20 MHz digital or analog oscilloscope equivalent or better Optical power meter OPM traceable to local national standards ST ST singlemode fiber patch ca...

Page 3: ...ad to BT1 on the source connect the negative lead to BT2 1 8 Set the laser source s CW MOD switch to the CW position 1 9 Set the laser source s power switch to either the 1310 nm or 1550 nm position Verify that the laser source LOW BATTERY LED is off and the SOURCE ACTIVE LED is on Step 2 Checking the Laser Drive Voltage Use the following procedure to check the drive voltage for the laser diodes a...

Page 4: ...LS 1310 1550 Laser Source 4 J2 J4 R14 R4 BT1 BT2 R35 R24 U3 R24 U3 R35 SW1 SW2 R14 R4 J4 J2 Front Back nu01f eps Figure 1 Jumpers and Calibration Measurement and Adjustment Points for SN 79130000 ...

Page 5: ... 15 mA at 9 2 dBm If the current reaches 60 mA before the power reading reaches 9 2 dBm the laser source requires repair 3 6 Turn off the laser source Disconnect the current meter Reinstall the jumper between J4 1 and J4 2 3 7 With the laser source set to 1310 nm verify that the power reading is 9 2 dBm 0 1 dB 3 8 Adjust the 1310 nm external potentiometer R24 counterclockwise to verify that the la...

Page 6: ...ce CW MOD switch to the MOD position 4 12 Verify that the power reading is 10 dBm 0 5 dBm Verify that the signal frequency is 2 kHz 100 Hz with a duty cycle of 50 5 4 13 Set the laser source CW MOD switch to CW Turn off the laser source Step 5 Reassembly and Final Test 5 1 Remove the singlemode cable from the laser source Put the dust cap onto the laser source optical ST connector 5 2 Apply insula...

Page 7: ... 2 V Step 3 Calibrating and Adjusting the 1310 nm Circuit Use the following procedure to make calibration adjustments to the 1310 nm laser diode circuit 3 1 Adjust the dc power supply to 9 V 0 25 V with the output current limited to 250 mA Connect the power supply s positive lead to BT1 connect the negative lead to BT1 3 2 Set up the DMM to measure dc volts less than 2 V Connect the negative probe...

Page 8: ...LS 1310 1550 Laser Source 8 C13 Z2 R20 Z5 R12 R31 R2 BT1 BT1 TP2 TP1 R29 R39 R40 nu02f eps Figure 2 Calibration Measurement and Adjustment Points for SN 79130000 ...

Page 9: ...st R40 clockwise until either the power reading is 9 2 dBm 0 1 dB or the current reaches 60 mA A typical current value is 15 mA at 9 2 dBm If the current reaches 60 mA before the power reading reaches 9 2 dBm the laser source requires repair 4 6 With the laser source set to 1550 nm verify that the power reading is 9 2 dBm 0 1 dB 4 7 Adjust the 1550 nm external potentiometer R31 counterclockwise to...

Page 10: ...Top case 625837 Bottom case 625829 Fiber optic cable assembly ST ST singlemode 662368 Fiber optic adapter ST ST singlemode 625928 9 V battery 614487 LS 1310 1550 Laser Source Instruction Sheet 687290 Table 3 Accessories Description Fluke Model Number Fiber optic cable assembly ST ST singlemode NF100SM Fiber optic cable assembly ST SC singlemode NF110SM Fiber optic cable assembly ST FC singlemode N...

Reviews: