Fluke LS-1310 Information Sheet Download Page 3

Service Information Sheet

Calibration Adjustments for Serial Numbers 

<

 79130000

3

Calibration Adjustments for Serial
Numbers < 79130000

(for later serial numbers, see page 6)

To ensure that the laser source performs to
specifications, calibrate it annually using the
procedure that follows.

Always perform the complete calibration
adjustment procedure.

Step 1:  Preparing for Calibration

Before you start the calibration procedures
described in this sheet, do the following (refer to
Figure 1 for locations of components on the circuit
board):

1-1.

 

Turn off the laser source.

1-2.

 

Remove the battery door and the 9 V battery.

1-3.

 

Remove the four Phillips screws from the back
of the source. Remove the top case.

1-4.

 

Unscrew the external retaining nuts from the
source’s power switch and 

CW/MOD

 switch.

1-5.

 

Carefully remove the laser source circuit board
from the bottom case. Take care to not strain
the fiber that connects the circuit board to the
ST connector mounted in the bottom case.

1-6.

 

Before applying power to the source, do the
following:

a.

 

Remove the jumpers from J2 and J4.

b.

 

Adjust the internal potentiometers R4 and
R14 to their full counterclockwise
positions. You will hear a faint click at the
full counterclockwise position.

c.

 

Adjust the external 

OUTPUT ADJUST

potentiometers (R24 and R35) to their full
clockwise positions. You will hear a faint
click at the full clockwise position.

1-7.

 

Adjust the dc power supply to 10 V ±0.25 V
with the output current limited to 250 mA.
Connect the power supply's positive lead to
BT1 on the source; connect the negative lead
to BT2.

1-8.

 

Set the laser source's 

CW/MOD

 switch to the

CW

 position.

1-9.

 

Set the laser source's power switch to either
the 1310 nm or 1550 nm position. Verify that
the laser source 

LOW BATTERY

 LED is off

and the 

SOURCE ACTIVE

 LED is on.

Step 2:  Checking the Laser Drive
Voltage

Use the following procedure to check the drive
voltage for the laser diodes and to verify that the
laser source's low battery detection circuit is
operating properly.

2-1.

 

Set up the DMM to measure dc voltage.
Connect the DMM to the laser drive voltage
between J2-6 and J2-4 (Gnd). Verify that the
voltage is between 3.2 V dc and 3.5 V dc.
Record this measurement for use in the next
step.

 

Note

 

To avoid overshooting the target voltage in the
next step, slow the adjustment rate on the
power supply as the voltage approaches 7 V.

2-2.

 

Decrease the power supply voltage just until
the laser source's 

LOW BATTERY

 LED lights.

Verify the following:

a.

 

The power supply voltage is between
5.4 V dc and 6.2 V dc.

b.

 

The DMM reading has changed less than
0.3 V dc from the reading you recorded in
step 1.

2-3.

 

Connect the DMM's positive lead to U3-5.
Verify the reading to be < 500 mV dc.

2-4.

 

Turn off the laser source and the power
supply. Remove the power supply leads from
the laser source's battery terminals.

2-5.

 

Reassemble the bottom case, as follows:

a.

 

Verify that the star washers are in place at
the bases of the laser source switches; then
carefully reinstall the PCA into the bottom
case. Ensure that the fiber connecting the
circuit board to the ST connector is not
pinched between the board and the case.

b.

 

Reinstall the external retaining nuts on the
laser source's power switch and 

CW/MOD

switch.

Summary of Contents for LS-1310

Page 1: ...r radiation and to prevent eye damage Never look directly into the source connector Though the laser radiation is invisible it can damage your eyes Do not modify the source Do not magnify or otherwise modify the laser output Use only approved connectors and adapters Do not use controls adjustments or procedures not documented or approved by Fluke Corporation Caution While servicing the laser sourc...

Page 2: ...ance tests and calibration adjustments Fiber optic power meter calibrated at 1310 nm and 1550 nm 0 10 V dc adjustable power supply with a minimum output of 500 mA and a current limit adjustment Fluke 187 DMM digital multimeter equivalent or better 20 MHz digital or analog oscilloscope equivalent or better Optical power meter OPM traceable to local national standards ST ST singlemode fiber patch ca...

Page 3: ...ad to BT1 on the source connect the negative lead to BT2 1 8 Set the laser source s CW MOD switch to the CW position 1 9 Set the laser source s power switch to either the 1310 nm or 1550 nm position Verify that the laser source LOW BATTERY LED is off and the SOURCE ACTIVE LED is on Step 2 Checking the Laser Drive Voltage Use the following procedure to check the drive voltage for the laser diodes a...

Page 4: ...LS 1310 1550 Laser Source 4 J2 J4 R14 R4 BT1 BT2 R35 R24 U3 R24 U3 R35 SW1 SW2 R14 R4 J4 J2 Front Back nu01f eps Figure 1 Jumpers and Calibration Measurement and Adjustment Points for SN 79130000 ...

Page 5: ... 15 mA at 9 2 dBm If the current reaches 60 mA before the power reading reaches 9 2 dBm the laser source requires repair 3 6 Turn off the laser source Disconnect the current meter Reinstall the jumper between J4 1 and J4 2 3 7 With the laser source set to 1310 nm verify that the power reading is 9 2 dBm 0 1 dB 3 8 Adjust the 1310 nm external potentiometer R24 counterclockwise to verify that the la...

Page 6: ...ce CW MOD switch to the MOD position 4 12 Verify that the power reading is 10 dBm 0 5 dBm Verify that the signal frequency is 2 kHz 100 Hz with a duty cycle of 50 5 4 13 Set the laser source CW MOD switch to CW Turn off the laser source Step 5 Reassembly and Final Test 5 1 Remove the singlemode cable from the laser source Put the dust cap onto the laser source optical ST connector 5 2 Apply insula...

Page 7: ... 2 V Step 3 Calibrating and Adjusting the 1310 nm Circuit Use the following procedure to make calibration adjustments to the 1310 nm laser diode circuit 3 1 Adjust the dc power supply to 9 V 0 25 V with the output current limited to 250 mA Connect the power supply s positive lead to BT1 connect the negative lead to BT1 3 2 Set up the DMM to measure dc volts less than 2 V Connect the negative probe...

Page 8: ...LS 1310 1550 Laser Source 8 C13 Z2 R20 Z5 R12 R31 R2 BT1 BT1 TP2 TP1 R29 R39 R40 nu02f eps Figure 2 Calibration Measurement and Adjustment Points for SN 79130000 ...

Page 9: ...st R40 clockwise until either the power reading is 9 2 dBm 0 1 dB or the current reaches 60 mA A typical current value is 15 mA at 9 2 dBm If the current reaches 60 mA before the power reading reaches 9 2 dBm the laser source requires repair 4 6 With the laser source set to 1550 nm verify that the power reading is 9 2 dBm 0 1 dB 4 7 Adjust the 1550 nm external potentiometer R31 counterclockwise to...

Page 10: ...Top case 625837 Bottom case 625829 Fiber optic cable assembly ST ST singlemode 662368 Fiber optic adapter ST ST singlemode 625928 9 V battery 614487 LS 1310 1550 Laser Source Instruction Sheet 687290 Table 3 Accessories Description Fluke Model Number Fiber optic cable assembly ST ST singlemode NF100SM Fiber optic cable assembly ST SC singlemode NF110SM Fiber optic cable assembly ST FC singlemode N...

Reviews: