![Fluke i6000s Flex Series Instruction Sheet Download Page 6](http://html1.mh-extra.com/html/fluke/i6000s-flex-series/i6000s-flex-series_instruction-sheet_2309915006.webp)
i6000s English Instruction Sheet
Page 6
Battery Installation
Warning
To prevent electric shock or personal injury:
•
Remove Probe from energized circuits before
opening battery door.
•
Never operate the unit without the battery cover
installed.
The Probe requires two AA IEC LR6 alkaline batteries for
operation. The battery compartment is accessed from the rear
end of the electronic enclosure.
The batteries must be replaced when the LED is lit continuously
or when it fails to light up. Ensure that the Probe is away from
any current carrying conductor and also that the output is
disconnected from other equipment.
To install the battery:
1. Use a coin or a similar tool to turn the battery lock (¼ turn)
until the dot aligns with the unlock symbol.
2. Remove the battery cover.
3. Install the batteries ensuring that correct polarity is
observed.
4. Replace the battery cover and turn the battery lock until the
dot aligns with the lock symbol.
Measuring Current
Warning
To prevent electric shock or personal injury:
•
Read Safety Instructions before operating this
product.
•
Ensure that the conductor to be tested is de-
energized.
To measure current:
1. Connect the output of the electronics to the input of an
oscilloscope or other data recording device.
Warning
To prevent electric shock or personal injury, the
flexible current Probes are not for use on
conductors with a potential of over 600 V.