14
Dynamic Characteristics
Bandwidth
DC to 50 MHz (-3 dB)
Frequency derating
(see Figure 8)
10 A at 10 MHz
Rise time, (10 % to 90 %)
tr < 7ns
Delay time
td < 25 ns
Overshoot
< 5 % of reading
General Data
Aperture dimensions
5 x 5 mm (0.2 in x 0.2 in)
Max primary conductor temp. 60
°
C (140
°
F)
Dielectric withstand
1350 Vrms / 50 Hz / 1 min
Working Voltage
CAT I
300 Vrms or dc, Pollution Degree 2
CAT II
150 Vrms or dc, Pollution Degree 2
Operating temperature
0
°
C to + 40
°
C (32
°
F to 104
°
F)
Storage temperature
-10
°
C to + 60
°
C (14
°
F to 140
°
F)
Maximum
2000 m (6600 ft)
Maximum relative humidity 80 %, 31
°
C (87
°
F)
Environment
indoor use only
Power supply
± 12 V, ± 0.5 V
Current consumption
at nominal
30 A , 550 mA
during demagnetization 1.3 A (for 6 sec)
Probe dimensions
Length
191.1 mm (7.53 in)
Width
28.9
mm
(1.14
in)
Summary of Contents for i50s