Wireless iFlex 

 

Before Calibration Adjustment 

Before Calibration Adjustment 

Before the Product calibration can be adjusted, you must go through the 
maintenance mode menu and enter your password. 

Maintenance Mode 

The Product maintenance mode can be used to set different parameters on the 
Product that include auto power off, backlight adjustment, and calibration. To use 
the maintenance mode: 

1.  With the Product off, push and hold 

2. Push 

. Keep 

 pushed until all the display segments are shown. 

3. Release 

 and 

The Product is now in maintenance mode.  

Password Entry 

To go to the calibration mode, push 

 until 

CAL

 is shown. You will need to 

enter a password to access calibration mode. 

To enter the password: 

1. Push 

 and the CAL count is shown. For example 

n002

2. Push 

 to show “

????

”. The first “

?

” flashes.  

3. Push 

 to change the flashing “

?

” to the first digit of your password 

(default: 1234). 

4. Push 

 to confirm your choice. The subsequent “

?

” flashes. 

5.  Do steps 3 and 4 again to enter the subsequent digits of the four-number 

password. 

6.  When all of the correct digits are entered, push 

 to confirm the input.  

If the correct password is entered, “

C-01

” is shown. If the incorrect password 

is entered, “

????

” is shown and the password must be entered correctly to 

go to the first calibration point, “

C-01

”.  

Change the Password 

Note 

If you change the password and then lose it, see the “Restore the 
Default Password” section. 

To change the password: 

1.  Do steps 1 through 5 in the “Password Entry” section.  
2. Before you push 

 to confirm your final input (step 6), push 

 to show “----” 

on the display. The first “-” flashes.  

3. Push 

 to change the first “-” to the first digit of your new password. 

4. Push 

 to confirm your choice. The next “-” flashes. 

5.  Repeat steps 3 and 4 to enter the subsequent digits of the new four-number 

password. 

6.  When the correct digits are entered, push 

 to confirm the input and change 

the password. If the Product has been calibrated, it will go to normal 
measurement mode, or it will show “

donE

”. 

Summary of Contents for a3001 FC

Page 1: ...2014 2014 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies a3001 FC Wireless iFlex Calibration...

Page 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Page 3: ...pment 5 Performance Tests 5 Test the Display 5 Backlight 6 Keypad Test 6 AC Current Test 6 Before Calibration Adjustment 9 Maintenance Mode 9 Password Entry 9 Change the Password 9 Restore the Default...

Page 4: ...a3001 FC Calibration Manual ii...

Page 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Required Equipment 5 3 Performance Tests 7 4 Calibration Adjustment 11 5 User Replaceable Parts 13...

Page 6: ...a3001 FC Calibration Manual iv...

Page 7: ...Figure Title Page 1 All Segments of the Display 5 2 Performance Test Connections for Simulated Voltages 7 3 Performance Test Connections for Applied Current 8 4 Calibration Password Reset 10 5 Batter...

Page 8: ...a3001 FC Calibration Manual vi...

Page 9: ...call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31 402...

Page 10: ...must be closed and locked before you operate the Product Replace the batteries when the low battery indicator shows to prevent incorrect measurements Have an approved technician repair the Product Use...

Page 11: ...ment Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation CAT IV Measurement Category IV is applicable to test...

Page 12: ...C or 28 C Add 0 1 X specified accuracy F 64 4 F or 82 4 F Safety IEC 61010 1 Pollution Degree 2 Electromagnetic Environment IEC 61236 1 Portable Electromagnetic Compatibility Radio Frequency Emission...

Page 13: ...prevent possible electrical shock fire or personal injury do not perform the performance test procedures unless the Product is fully assembled The performance tests verify the full operation of the Pr...

Page 14: ...you have the necessary equipment See Table 2 2 Make sure the Product battery is good and replace it if necessary See Battery Replacement 3 Warm up the Calibrator as necessary Refer to its specificati...

Page 15: ...Simulate 67 5 mV 45 Hz 1 2420 2580 Single Turn Loop 10 A 500 Hz 0 1 9 2 10 8 Simulate 180 mV 500 Hz 0 1 581 5 618 5 Simulate 300 mV 500 Hz 1 965 1035 Simulate 750 mV 500 Hz 1 2420 2580 50 Turn Coil 4...

Page 16: ...LO TC 1000V RMS MAX 20V RMS MAX 20V PK MAX 1VPK MAX 20VPK MAX NORMAL AUX TRIG OUT SCO PE V RTD A SENSE AUXV 150VPK MAX 20VPK MAX 22 50 Turn Current Coil a3001 FC Conductor hca009 eps Figure 3 Performa...

Page 17: ...ush to confirm your choice The subsequent flashes 5 Do steps 3 and 4 again to enter the subsequent digits of the four number password 6 When all of the correct digits are entered push to confirm the i...

Page 18: ...ps screwdriver remove the bottom case screws Two of the screws are inside of the battery door 3 Keep the pca in the top case 4 Apply 3 0 V across the battery contacts on the pca Note the polarity that...

Page 19: ...push wait until the calibration step number advances before you change the calibrator source Some adjustment steps can take several seconds to execute before the Product goes to the subsequent step S...

Page 20: ...t or to equipment under test Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period Be sure that the battery polarity is correct to prevent batt...

Page 21: ...r replaceable parts are shown in Table 5 Table 5 User Replaceable Parts Fluke Part Number Description Qty 4130305 FLK 3000 2003 DOOR BATTERY 1 3676410 FLUKE I2500 10 IFLEX 2500A PROBE 10IN 1 1881997 T...

Page 22: ...a3001 FC Calibration Manual 14...

Reviews: