Table 3. Performance Tests
Test Connection
Calibrator
Output Value
Res. Spec.
UUT Reading Limit
Low High
Single-Turn Loop
0 A,0 Hz
0.1
3.0 %
±
5 digits
0
0.5
Single-Turn Loop
10 A,45 Hz
0.1
9.2
10.8
50-Turn Coil
12 A,45 Hz
0.1
581.5
618.5
50-Turn Coil
20 A,45 Hz
1
965
1035
Simulate
67.5 mV,45 Hz
1
2420
2580
Single-Turn Loop
10 A,500 Hz
0.1
9.2
10.8
Simulate
180 mV,500 Hz
0.1
581.5
618.5
Simulate
300 mV,500 Hz
1
965
1035
Simulate
750 mV,500 Hz
1
2420
2580
50-Turn Coil
45 Hz,0.4 A
0.1
0.5%
±
5 digits
44.3
45.7
50-Turn Coil
500 Hz,0.325 A
0.1
497.0
503.0