Replacing the Line-Power Fuses
XW
Warning
To prevent fire hazard or damage to the instrument, verify that
the correct fuses are installed for the selected line-voltage
setting. See Table 2-4 for the correct fuse ratings.
The line-power fuses are located on the power block on the rear panel. The selected line-
voltage (115 or 230) shows through a small window toward the bottom of the block. See
Figure 2-1. This instrument has dual fusing with fuses in both the line and neutral
connections. When replacing a blown fuse, replace both fuses to avoid a stressed fuse and
subsequent power interruption. Before trying to access and replace the fuses, verify that
the replacement fuses are appropriate for the selected voltage.
To check or replace the fuses, refer to Figure 2-1, and proceed as follows:
1.
Disconnect the Instrument from line power.
2.
Remove the fuse compartment by inserting a screwdriver blade in the tab located at
the left side of the fuse compartment. Gently pry until the compartment can be easily
removed.
3.
Pull the fuses from the compartment for replacement or inspection.
4.
Install good fuses with the correct ratings. See Table 2-4.
5.
Reinstall the fuse compartment by pushing it back into place until the tab locks.
Table 2-4. Line-Power Fuse
Line Power
Fuse Action
Fuse Rating
IEC 127
Fluke Part No.
Manufacturer
and Type No.
115 V AC
TH
Time Delay HBC
10 A @ 250 V
2650727
(Quantity 2)
Littelfuse 215010
230 V AC
TH
Time Delay HBC
5 A @ 250 V
2650730
(Quantity 2)
Littelfuse 215005
Selecting Line Voltage
The line-voltage selector is located on the power block on the rear panel. The selected
line-voltage (115 or 230) shows through a small inspection window toward the bottom of
the block. See Figure 2-1. Use Table 2-5 to verify the selection before trying to change it.
Use the following procedure to change the line voltage setting:
1.
Disconnect the Instrument from line power.
2.
Remove the fuse compartment as described earlier. (See
Replacing the Line-Power
Fuse
.)
3.
Remove the line-voltage selector by griping its indicator tab with a pair of long-nose
pliers and pulling it straight out of its connector.
4.
Rotate the line-voltage selector to show the desired voltage, and reinsert into the
power block.
5.
Before inserting the fuse compartment back into the power block, inspect and verify
that both fuses are appropriate for the selected voltage.
6.
Insert the fuse compartment into the power block, and press firmly to lock its tab.
Summary of Contents for 9640A Series
Page 8: ......
Page 16: ...9640A Instruction Manual viii...
Page 22: ...9640A Instruction Manual 1 2...
Page 40: ...9640A Instruction Manual 2 2...
Page 108: ...9640A Instruction Manual 4A 2...
Page 138: ...9640A Instruction Manual 4C 2...
Page 164: ...9640A Instruction Manual 4D 2...
Page 168: ...9640A Instruction Manual 4D 6...
Page 170: ...9640A Instruction Manual 4E 2...
Page 174: ...9640A Instruction Manual 4E 6...
Page 176: ...9640A Instruction Manual 4F 2...
Page 184: ...9640A Instruction Manual 4F 10...
Page 186: ...9640A Instruction Manual 4G 2...
Page 190: ...9640A Instruction Manual 4G 6...
Page 192: ...9640A Instruction Manual 5 2...
Page 272: ...9640A Instruction Manual 5 82...
Page 274: ...9640A Instruction Manual 6 2...
Page 284: ...9640A Instruction Manual 7 2...
Page 310: ...9640A Instruction Manual 8 2...
Page 316: ...9640A Instruction Manual 8 8...
Page 320: ...9640A Instruction Manual A 4...