Table 3. Controller Communications Commands
Command
Description
Command
Format
1
Command
Example
2
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint]
s
set: 999.99 {C or F}
set:
150.00 C
-
Set current set-point
s[etpoint]=n
s=350
-
-
Instrument Range
Read temperature
t[emperature]
t
t{h or c}: 9999.99
{
C or F}
th: 55.66 C
-
Read temperature units
u[nits]
u
u: x
u: C
Set temperature units:
u[nits]=c/f
-
-
-
C or F
Set temperature units to
Celsius
u[nits]=c
u=c
-
-
-
Set temperature units to
Fahrenheit
u[nits]=f
u=f
-
-
-
Read scan mode
sc[an]
sc
sc:
{
ON or OFF}
sc:ON
-
Set scan mode
sc[an]=on/off
sc=on
-
-
ON or OFF
Read scan rate
sr[ate]
sr
srat: 99.9
{
C or F}/min
srat:12.4C/min
-
Set scan rate
sr[ate]=n
sr=1.1
-
-
.1 to 99.9
Secondary Menu
Read proportional band
setting
pr[opband]
pr
pb: 999.9
pb: 15.9
-
Set proportional band to
n
pr[opband]=n
pr=8.83
-
-
Depends on
Configuration
Read heater power (duty
cycle)
po[wer]
po
po: 999.9
po: 1.3
-
Configuration Menu
Operating Parameters Menu
Read High Limit
hl
hl
hl:999
hl:350
Set High Limit
hl=
n
hl=350
Cold 25 to 126,
Hot 50 to 350
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
-
Set serial sampling setting to
n
seconds
sa[mple]=n
sa=0
-
-
0 to 999
Set serial duplex mode:
du[plex]=f[ull]/
h[alf]
-
-
-
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
-
-
-
Set serial duplex mode to half
du[plex]=h[alf]
du=h
-
-
-
Set serial linefeed mode:
lf[eed]=on/of[f]
-
-
-
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
-
-
-
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
-
-
-