87V EX
Calibration Manual
18
Table 5. Accuracy Tests (cont.)
Step
Test
Function
Range
5500A Output
Display Reading
23 600
Ω
330
Ω
( Use 2 wire Comp)
1
329.1 to 330.9
24 6
k
Ω
3.3
k
Ω
(Use 2 wire Comp)
1
3.292 to 3.308
25 60
k
Ω
33
k
Ω
32.92 to 33.08
26 600
k
Ω
330
k
Ω
327.9 to 332.1
27 6
M
Ω
3.3
M
Ω
3.279 to 3.321
28
e
Ohms
50 M
Ω
30
M
Ω
29.67 to 30.33
29
60 nS
Open input
- 0.30 to 0.30
30
nS
Conductance 60 nS
100 M
Ω
9.60 to 10.40
31
G
Diode
6 V
3.0 V dc
2.939 to 3.061
32
?
AC Amps
6 A
3.0 A, 60 Hz
2.968 to 3.032
33
A
DC Amps
6 A
3.0 A
2.990 to 3.010
33B
A
DC Amps
10A
6
10A
9.96 to 10.04
34
60 mA
33 mA, 60 Hz
32.65 to 33.35
35
m
?
AC Milliamps 400 mA
330 mA, 60 Hz
326.5 to 333.5
36
60 mA
33 mA
32.89 to 33.11
37
m
[
DC Milliamp
400 mA
330 mA
329.1 to 330.9
38
600
µ
A 330
µ
A, 60 Hz
326.5 to 333.5
39
U
?
AC
Microamps
6000
µ
A 3300
µ
A, 60 Hz
3265 to 3335
40
600
µ
A 330
µ
A
328.9 to 331.1
41
U
[
DC
Microamps
6000
µ
A 3300
µ
A
3291 to 3309
42
10 nf
Open input
2
0.21 to 0.31
43
100 nf
5 nf
5
04.7 to 05.3
44
E
Capacitance
100
µ
f 9.5
µ
f
09.2 to 09.8
45
1000 V
400 V, 400 Hz
372 to 408
46
K
Low Pass
Filter
1000 V
400 V, 800 Hz
4
226 to 340
4