background image

 

F1.eps 

Figure 1. Initial Case Temperature Above Ambient vs Meter 

Reading Above Ambient 

Operation 

Use the following procedure to operate the 80T-150UA probe: 

1.

 

Connect the banana plugs on the 80T-150UA to the input terminals 
of a high impedance DMM. Observe polarity. 

2.

 

Select a dc voltage range that will provide at least 1 mV resolution  
(1 mV/degree) and a full scale readout that will encompass the 
expected temperature. The 2 V range of a  
3 1/2-digit DMM is adequate. Ignore readings of less than 1

°

 when   

a more sensitive DMM is used. 

3.

 

Set the 80T-150UA power switch to 

°

C or 

°

F, and energize the 

DMM. 

4.

 

Firmly touch the probe tip to the surface to be measured, or expose 
it to a liquid or gas. The DMM will display the temperature in 
degrees. Vary the probe angle and pressure when measuring solid 
surface temperatures. The highest stabilized reading will be the 
most accurate. (See the following measuring technique.) 

W

Caution 

The force exerted on the probe tip should not exceed 15 
pounds. 

Measuring Technique 

Here are some suggestions for improving the accuracy of your 
temperature measurements: 

1.

 

When measuring higher than ambient temperatures, adjust the 
connection between the probe and the surface until you get the 
highest temperature reading. 

2.

 

When measuring lower than ambient temperatures, adjust the 
connection between the probe and the surface until you get the 
lowest temperature reading. 

3.

 

When measuring near ambient temperatures, make the reading 
when the multimeter readout is most stable. 

General Maintenance 

Access Information 

The battery and the calibration pots are located on the interior of the 
temperature-to-voltage converter assembly. Access to these locations is 
accomplished by removing the screw from the bottom side of the 
assembly and removing the top of the plastic box. 

Battery Condition Test 

1.

 

Set the power switch to the OFF position. 

2.

 

Connect the 80T-150UA to the DMM. 

3.

 

Set the DMM to the 200 or 400 mV dc range. 

4.

 

Read the battery test voltage on the DMM or test instrument. For 
DMMs with 10 M

Ω

 input impedance, a minimum reading of 100 mV 

is acceptable. For test instruments with 1 M

Ω

 input impedances, 

such as ScopeMeter

®

, a minimum reading of 75 mV is acceptable, 

With these minimum readings, approximately 100 hours of battery 
life remain. 

Reviews: