Operating Instructions

XW

 Warning

To prevent possible electrical shock, fire, or personal injury:

• When using the Clamp, ensure that your fingers are 

behind the protective barrier as shown in Table 1.

• Do not use the Clamp if any part, including the lead and 

connector(s), appears to be damaged or if a malfunction 
of the instrument is suspected.

• Before using the clamp make sure there is no dust or any 

small debris on the closing faces of the jaw.

• Close the jaws gently and avoid closing them with high-

impact. The force of the impact can affect the accuracy of 
the measurement.

Turn on the Clamp

Push 

+

 for 2 seconds to turn on/off the clamp. 

Green LED indicates the Clamp is ON and the battery is ok.

Orange LED indicates the battery is lower than 2.2 V.

Red LED indicates the battery is lower than 2 V. Replace the 
battery before doing a measurement.

P

Conforms to European Union directives.

Conforms to relevant Australian Safety and EMC 
standards.

Conforms to relevant South Korean EMC Standards.

Measurement Category III is applicable to test and 
measuring circuits connected to the distribution part of 
the building’s low-voltage MAINS installation.

Measurement Category IV is applicable to test and 
measuring circuits connected at the source of the 
building’s low-voltage MAINS installation.

~

This product complies with the WEEE Directive marking 
requirements. The affixed label indicates that you must 
not discard this electrical/electronic product in domestic 
household waste. Product Category: With reference to 
the equipment types in the WEEE Directive Annex I, this 
product is classed as category 9 "Monitoring and 
Control Instrumentation" product. Do not dispose of this 
product as unsorted municipal waste.

Table 2. Symbols (Continued)

Symbol

Definition

Summary of Contents for 80i-2010s

Page 1: ...l for use in inverters switch mode power supplies industrial controllers automotive diagnostics and other applications requiring current measurements and or waveform analysis Table 1 shows the Clamp How to Contact Fluke To contact Fluke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 3...

Page 2: ...ersonal protective equipment approved rubber gloves face protection and flame resistant clothes to prevent shock and arc blast injury where hazardous live conductors are exposed Do not use the Product around explosive gas vapor or in damp or wet environments Examine the case before you use the Product Look for cracks or missing plastic Carefully look at the insulation around the terminals Table 1 ...

Page 3: ...ates incorrectly The battery door must be closed and locked before you operate the Product Remove all probes test leads and accessories before the battery door is opened Do not work alone Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product probe or accessory Do not touch voltages 30 V ac rms 42 V ac peak or 60 V dc Have an approved technician rep...

Page 4: ...tery is lower than 2 V Replace the battery before doing a measurement P Conforms to European Union directives Conforms to relevant Australian Safety and EMC standards Conforms to relevant South Korean EMC Standards Measurement Category III is applicable to test and measuring circuits connected to the distribution part of the building s low voltage MAINS installation Measurement Category IV is appl...

Page 5: ...t indicates that the current flow is in the direction shown by the arrow on the clamp Maintenance Before each use inspect the Clamp Look for cracks or missing portions of the housing and output cable insulating cover and for loose or weakened components Cleaning and Storage Periodically wipe the case with a damp cloth and detergent Do not use abrasives or solvents Do not immerse the Clamp in liqui...

Page 6: ... cable with BNC male connector Maximum Conductor Diameter 52 0 mm Maximum Jaw Opening 81 5 mm Output to Zero Auto adjustment with zero button Temperature Operating 10 C to 50 C Storage battery removed 30 C to 60 C Operating Humidity 90 RH at 10 C to 30 C 75 RH at 30 C to 40 C 45 RH at 40 C to 50 C Altitude Operating 2000 m Storage 12 000 m Demagnetize Clamp Open and gently close the Clamp jaws sev...

Page 7: ...ISPR 11 can occur when the equipment is connected to a test object Korea KCC Class A Equipment Industrial Broadcasting Communication Equipment Class A Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it This equipment is intended for use in business environments and not to be used in homes Electrical Specifications All accuraci...

Page 8: ...ion then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since some states or countries do not allow the exclusion or limitation o...

Reviews: