![Fluke 78 User Manual Download Page 59](http://html.mh-extra.com/html/fluke/78/78_user-manual_2309743059.webp)
Specifications
47
Function
Range
Resolution
Accuracy
Burden Voltage (Typical)
AC Volts*
(45 Hz to 1 kHz)
4.000 V
40.00 V
300.0 V
300 V
0.001 V
0.01 V
0.1 V
1 V
±
(2.5% + 2)
±
(2.5% + 2)
±
(2.5% + 2)
±
(2.5% + 2)
±
1.5dB typical
N/A
DC Volts*
400.0 mV
4000 mV
4.000 V
40.00 V
300.0 V
300 V
0.1 mV
1 mV
0.001 V
0.01 V
0.1 V
1 V
±
(0.3% + 5)
±
(0.3% + 1)
±
(0.3% + 1)
±
(0.3% + 1)
±
(0.3% + 1)
±
(0.3% + 1)
N/A
Resistance
400.0
Ω
4.000 k
Ω
40.00 k
Ω
400.0 k
Ω
4.000 M
Ω
40.00 M
Ω
0.1
Ω
0.001 k
Ω
00.01 k
Ω
000.1 k
Ω
0.001 M
Ω
0.01 M
Ω
±
(0.5% + 2)
±
(0.5% + 1)
±
(0.5% + 1)
±
(0.5% + 1)
±
(0.5% + 1)
±
(1% +3)
N/A
Continuity
400.0
Ω
0.1
Ω
Beeper on @ <30
Ω
for
short of 1 ms of longer
Open circuit voltage <1.5 V
Diode Test
2.500 V
0.001 V
±
2% typical
Open circuit voltage <3.3 V
AC Current
(45 Hz to 1 kHz)
10.00 A**
0.01 A
±
(2.5% + 2)
0.03 V/A
DC Current
4.000 A
10.00 A**
0.001 A
0.01 A
±
(1.0% + 5)
±
(1.0% + 2)
0.03 V/A
0.03 V/A
*
Input Impedance: 10 M
Ω
(nominal), <150 pF>
** 10 A continuous, 20 A Overload for 30 seconds maximum.