Digital Multimeter
General Specifications
17
Function Overload
Protection
[1]
Input Impedance
(Nominal)
Common Mode Rejection Ratio
(1 k
Ω
Unbalanced)
Normal Mode Rejection
Volts AC
1000 V
>10 M
Ω
<100 pF
>60 dB @ dc, 50 Hz or 60 Hz
Volts DC
1000 V
>10 M
Ω
<100 pF
>120 dB @ dc, 50 Hz or 60 Hz
>60 dB @ 50 Hz or 60 Hz
mV
1000
V
[2]
>10
M
Ω
<100 pF
>120 dB @ dc, 50 Hz or 60 Hz
>60 dB @ 50 Hz or 60 Hz
Open Circuit Test
Voltage
Full Scale Voltage To:
6.0 M
Ω
50
M
Ω
Short Circuit Current
Ohms/Capacitance
1000
V
[2]
<8.0 V dc
<660 mV dc
<4.6 V dc
<1.1 mA
Continuity/Diode test
1000 V
[2]
<8.0 V dc
2.4 V dc
<1.1 mA
[1] 10
7
V-Hz maximum.
[2]
For circuits <0.3 A short circuit. 660 V for high energy circuits.
Function
Overload
Protection
Overload
mA
Fused, 440 mA, 1000 V FAST Fuse
600 mA overload for 2 minutes maximum, 10
minutes rest.
A
Fused, 11 A, 1000 V FAST Fuse
20 A overload for 30 seconds maximum, 10
minutes rest.
Summary of Contents for 77 Series IV