![Fluke 729 Pro User Manual Download Page 23](http://html1.mh-extra.com/html/fluke/729-pro/729-pro_user-manual_2310413023.webp)
Automatic Pressure Calibrator
Setup Menu
17
4. To let the water out, turn the knob counterclockwise on the side of the Product.
5. Push
D
to repeat the drain operation.
6. When finished with water drain, push in on the fitting to release the hose.
7. Turn the knob clockwise after you drain the water.
Note
Make sure this knob is tight. Otherwise, a leak can occur.
Figure 4. Drain Water from Product
Keypad Test
Use this screen to check the Product buttons. Push any button and verify the on-screen
response.
For the power button, push and hold for 2 seconds.
Exhaust
The Exhaust function releases all pressure inside the Product including the pressure
reservoir. This reservoir keeps the pressure stable and provides pressure if only small
pressure steps are necessary.
1. Push
A
for
Exhaust
.
2. Push
A
for
Abort
when done.