Percent Error Setup
Press and hold
a
. After 3 seconds the set icon and 0% appears on the lower
display. Use
W
and
X
to adjust the 0% point for the Percent Error calculation,
then press
D
(ENTER) key to confirm selection. Press
a
. 100% is displayed
on the lower display. Use
W
and
X
to adjust the 100% point for the Percent Error
calculation. Press
D
(ENTER) to confirm the selection and exit.
Calibrating a P/I Transmitter
To calibrate a P/I (pressure to current) transmitter:
1.
Open the pump vent and zero the Calibrator before applying pressure.
Repeat frequently to ensure accuracy.
2.
Apply a pressure to the transmitter and measure the transmitter’s current
loop output. OL (overload) is displayed at full scale.
3.
Connect the Calibrator to the transmitter as shown in Figure 4 or Figure 5.
The Calibrator recognizes only the pressure module if both types of
connection are in place.