Table 2. Specifications
1
(% of Maximum Rated Pressure)
Model
Maximum
Rated
Pressure
2
(Range)
Type
Isolated or
Nonisolated
Reference
Uncertainty
(23
°
±
3
°
C)
Stability
(1 Year)
Temp
(0 to 50
°
C)
Total
Uncertainty
3
Burst pressure: 3X maximum rated pressure, including common mode pressure (700P29Ex and 700P09Ex: 2x).
Specifications reflect a confidence interval of 95%.
700P01Ex
0 to 10 in H
2
O
0 to 2.49 kPa
0 to 0.02 bar
Differential
5
Hi:
Nonisolated
Low: Nonisolated
0.200 %
0.050 %
0.050 %
0.300 %
700P24Ex
0 to 15.000 psi
0 to 100.00 kPa
0 to 1.0000 bar
Differential
5
Hi:
Isolated
Low: Nonisolated
0.025 %
0.010 %
0.015 %
0.050 %
700P05Ex
0 to 30.000 psi
0 to 200.00 kPa
0 to 2.0000 bar
Gage Isolated
0.025 %
0.010 %
0.015 %
0.050 %
700P06Ex
0 to 100.00 psi
0 to 700.00 kPa
0 to 7.0000 bar
Gage Isolated
0.025 %
0.010 %
0.015 %
0.050 %
700P27Ex
0 to 300.00 psi
0 to 2000.00 kPa
0 to 20.000 bar
Gage Isolated
0.025 %
0.010 %
0.015 %
0.050 %
700P09Ex
0 to 1500.0 psi
0 to 10000.0 kPa
0 to 100.000 bar
Gage Isolated
0.025 %
0.010 %
0.015 %
0.050 %
700P29Ex
4
0 to 3000 psi
0 to 20680 kPa
0 to 207 bar
Gage,
High Pressure
Isolated
0.050 %
0.010 %
0.020 %
0.080 %
700PA4Ex
0 to 15.000 psi
0 to 100.00 kPa
0 to 1000.0 mbar
Absolute
Isolated
0.050 %
0.010 %
0.010 %
0.070 %
1.
Use of pressure zero function is required to achieve these specifications.
2.
Available pressure units are determined by the calibrator being used.
3.
Accuracy specifications apply for 1 year for 0 to 100% of full span from 0 to 50
°
C. Typical uncertainty is 1% of full span from -10
°
C to 0
°
C.
Maximum altitude: 2000 m. Humidity range: 0 to 80 %.
4.
Only use with Group 2 Fluids compatible with Hastelloy C276 and type 316 Stainless Steel.
5.
The maximum line (common mode) pressure shall not exceed the maximum rated pressure.
Table 3. Entity Parameters
Vmax, Ui
Imax, Ii
Pi
Ci
Li
8.7 V
598 mA
1.2 W
5.72
µ
F 0
mH