Fluke 700P01Ex Instruction Sheet Download Page 3

 

 

15 P

SID

/G

PRE

SSU

RE M

ODU

LE

700P

04

1 ba

r

100

 kPa

RAN

GE

 

Figure 1. 

Avoiding Overpressure Damage 

Applying pressure in excess of the burst pressure specified 
on the pressure modules can destroy the pressure modules.  

W

Caution 

Exceeding the maximum rated pressure 
may affect total uncertainty. If you 
suspect the module has been over 
pressured, use the performance test to 
check it.  

Avoiding Corrosion Damage 

To avoid corrosion damage, use only with specified media as 
shown below: 

 

Isolated: any medium that is compatible with type 316 
stainless steel. 

 

Non-Isolated: dry, non-corrosive gasses only. 

 

700P29Ex: use only with a medium that is compatible 
with Hastelloy C276 and type 316 Stainless Steel. 

Recommended Measurement Technique 

For best results, the modules should be pressurized to full 
scale and then vented to zero pressure (atmosphere) prior to 
zeroing and making measurements.  

Zeroing Gage and Differential Modules 

1.

 

Connect the pressure module to the intrinsically safe 
calibrator and select the measure pressure function. 

2.

 

Position the module in the same orientation for which it 
will be used. 

3.

 

Vent both high and low side measurement ports to 
atmosphere. 

4.

 

Press the [ZERO] key. 

 

Zeroing Absolute Modules 

1.

 

Connect the pressure module to the intrinsically safe 
calibrator and select the measure pressure function. 

2.

 

Apply a vacuum to achieve a pressure below the rated 
resolution of the pressure module being zeroed. 

3.

 

Press the [ZERO] key, and enter 0.0 as the applied 
pressure. 

Alternative procedure if a local precision barometer is 
available. Do not use the weather service or airport reports. 

1.

 

Connect the pressure module to the intrinsically safe 
calibrator and select the measure pressure function. 

2.

 

Press the [ZERO] key. 

3.

 

Enter the value from the precision barometer. 

Note 

Low range pressure modules may be sensitive to 
gravity. For best results, pressure modules 30 psi 
and below should be held at the same physical 
orientation from the time they are zeroed until the 
measurement is complete. 

Pressure Calibration Kit 

The Fluke-700PCK Pressure Calibration Kit makes it 
possible to calibrate pressure modules at ambient 
temperature with a precision pressure calibrator better than 
the module specification. A PC running Windows

®

 software is 

required. The kit is an optional accessory available from your 
distributor or Fluke. 

Performance Test 

If you need to check that the pressure module meets its total 
uncertainty specification, use a dead weight tester or suitable 
pressure calibrator. Proceed as follows to verify that a 
pressure module is operating within specification: 
1.

 

Read the pressure value with no externally applied 
pressure to make sure the 0% of scale is correct. When 
reading the pressure, press the [ZERO]  key to remove 
any zero offset. 

2.

 

Connect the pressure module to a precision pressure 
source. 

3.

 

Zero as described earlier in the appropriate “Zeroing” 
sections. 

4.

 

Set the precision pressure source to 20% of the 
pressure module’s maximum rated pressure. 

5.

 

Make sure the reading agrees with precision pressure 
source value within the total uncertainty specification in 
Table 2. 

6.

 

Set the precision pressure source to 40, 60, 80, and 
100%. Then reverse the order from 100, 80, 60, 40, and 
20% of maximum rated pressure. Repeat step 5 at each 
test point. 

7.

 

If temperature sensitivity is of concern, repeat steps 1 
through 5 at various controlled temperatures. 

 

Cleaning 

W

Caution 

 

Measuring substances that 
leave residue in the sensor 
may cause permanent 
damage to the Pressure 
Module. 

 

Periodically wipe the Pressure 
Module with a damp cloth and 
mild detergent. Do not use 
abrasives or solvents to clean 
the Pressure Module.

Reviews: