Using Error Mode (cont.)
Keys
Action
E
Returns to the previous reference
value.
+
, then
E
Establishes a new reference.
A keypad entry, then
E
Establishes a new reference.
N
Establishes the present output as a
new reference.
Y
Sets the Calibrator to ten times the
reference value and establishes a new
reference.
Z
Sets the Calibrator to one-tenth the
reference value and establishes a new
reference.
o
Identifies the present output as a zero-
scale endpoint for scaling, and
establishes 0.0 as the new reference.
S
Identifies the present output as a full-
scale endpoint for scaling and causes
the display to show scale error.
r
Returns to the power-up state.
The “Setup Menus”
softkey
Opens the setup menu.