20
Front Panel Keys (cont.)
0
-
9
(Numeric Keypad)
Used to enter the digits of the output amplitude
and frequency, as well as other data such as the
time and date. The proper sequence for entering a
value is to enter the digits, then any necessary
multipliers, an output function key, then
E
.
For example, to obtain an output of 20 mV, you
would press the following sequence of keys:
2 0 m V E
. The
.
key is
available for decimal places.
+
If the output function is dc voltage, current, ac
voltage entered in dBm, or a wideband output
entered in dBm, pressing
+
E
toggles the
polarity of the output. If the output function is ac
voltage or current, pressing
+
E
changes
the output to dc.