5500A
Service Manual
6-58
2.
Using the BNC cable, connect the SCOPE connector to the BNC(f) connector
attached to the nominal resistance values indicated in Table 6-45. The 600 K
Ω
nominal value can be achieved by connecting the 1.5 M
Ω
and 1 M
Ω
resistors in
parallel.
3.
Allow the Calibrator Mainframe reading to stabilize, then record the Calibrator
Mainframe resistance reading for each nominal value listed in Table 6-45. Compare
the Calibrator Mainframe resistance readings to the actual resistance values and the
tolerance column of Table 6-45.
Table 6-45. MeasZ Resistance Verification
Calibrator
Mainframe
MeasZ
Range
Nominal
Resistance
Value
Calibrator
Mainframe
Resistance
Reading
Actual
Resistance
Value
Tolerance
res 50
Ω
40
Ω
0.04
Ω
res 50
Ω
50
Ω
0.05
Ω
res 50
Ω
60
Ω
0.06
Ω
res 1M
Ω
600 k
Ω
600
Ω
res 1M
Ω
1 M
Ω
1 k
Ω
res 1M
Ω
1.5 M
Ω
1.5 k
Ω
6-72. MeasZ Capacitance Verification
The MeasZ capacitance function is verified by measuring capacitors of known values.
The measurement value is then compared to the capacitor actual value.
The capacitors must make a solid connection to a BNC(f) to enable a connection to the
end of the BNC cable supplied with the SC600. Due to the small capacitance values, care
must be taken to know the actual capacitance at this BNC(f) connector. The capacitance
values must be determined at a 10 MHz oscillator frequency. Fluke uses an HP 4192A
Impedance Analyzer at 10 MHz to determine the actual capacitance values.
This procedure uses the following equipment:
•
Adapters and capacitors to achieve 5 pF, 29 pF, 49 pF nominal values at the end of
BNC(f) connector
•
BNC cable supplied with the SC600
Refer to Figure 6-17 for the proper setup connections.
Set the Calibrator Mainframe to SCOPE mode, with the MeasZ menu on the display.
Then follow these steps to verify the MeasZ capacitance function.
1.
Set the Calibrator Mainframe MeasZ capacitance range to cap. (The blue softkey
under MEASURE toggles the MeasZ ranges).
2.
Connect the BNC cable to the Calibrator Mainframe SCOPE connector, but do not
connect any thing to the end of this cable.
3.
Allow the Calibrator Mainframe reading to stabilize, then press the SET OFFSET
blue softkey to zero the capacitance reading.
4.
Connect the end of the BNC cable to the BNC(f) connector attached to the nominal
capacitor values indicated in Table 6-46.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......