Programming a Test Voltage
To set a test voltage in between the pre-set test voltages proceed as follows:
1.
With the Meter turned on, press the
H
button to select
TEST VOLTAGE
.
00k
0
0M
G
V
0G
T
00G
G
T
M
k
00M
M
500V
Test Voltage Display
ASW05F.EPS
2.
Press the
V
or
d
button to scroll through the pre-set test voltage options
(250 V, 500 V, 1000 V, 2500 V, and 5000 V). Select the voltage closest to
the level required.
The selected test voltage appears in the upper-right of the display.
3.
Press the
e
button.
TV=
xxxxV appears flashing in the lower-left of the
display.
4.
Press the
V
and
d
buttons to increment and decrement the voltage. When
the correct voltage level shows, do not press the
e
button. Doing so will
return the test voltage to the next lowest last pre-set voltage selection.
Instead press the
H
button to go to the next function.
Note
The test voltage can be up to 10 % higher than the test voltage you
select.