12 Multimeter
TESTING CONTINUITY AND MEASURING RESISTANCE
9
TESTING CONTINUITY AND MEASURING RESISTANCE
1. Insert the test leads in the jacks, and turn off power to the circuit under test.
External voltage across the components causes invalid readings.
2. Put the slide-switch in the continuity/ohms position (Figure 4).
To toggle between the continuity/diode and ohms functions, press [
g
].
3. Touch the probes to the test points.
4. In ohms, read the resistance on the display.
In continuity test, the beeper sounds continuously if continuity exists (resistance
< 25
Ω
.). Opens and shorts longer than 250 µs are detected. On the Fluke 12,
short-to-open and open-to-short transitions can be captured and visually
displayed. See “Capturing Continuity Intermittents”.
If the meter detects a voltage greater in magnitude than about 4.5 V and the meter is
not in the manual range mode, the meter automatically changes to the voltage
measurement function. (See “CHEK AND HOW TO USE IT”.)
Summary of Contents for 12
Page 2: ......
Page 7: ...12 Multimeter DISPLAY 3 DISPLAY Figure 1 Display ...
Page 8: ...12 Multimeter Users Manual 4 OPERATING FEATURES Figure 2 Operating Features ...
Page 12: ...12 Multimeter Users Manual 8 Figure 3 Measuring Voltage ...
Page 14: ...12 Multimeter Users Manual 10 Figure 4 Continuity and Resistance Measurements ...
Page 16: ...12 Multimeter Users Manual 12 Figure 5 Testing Diodes ...
Page 20: ...12 Multimeter Users Manual 16 Figure 6 Measuring Capacitance ...
Page 22: ...12 Multimeter Users Manual 18 Figure 7 Displaying Minimum and Maximum Reading ...
Page 24: ...12 Multimeter Users Manual 20 Figure 8 Maximum and Minimum Reading with Elapsed Time ...
Page 26: ...12 Multimeter Users Manual 22 Figure 9 Open to Short and Short to Open Transitions ...
Page 32: ...12 Multimeter Users Manual 28 ...