05-443/05-444
Users Manual
14
Model 05-444
Typical response and set points for the Product with internal energy-
compensated LND 71412 G-M detector is as follows:
•
Operating Voltage: 550 V dc
•
Threshold: 100 mV dc
•
Calibration Constant: 150 cpm/mR/hr
•
Dead Time Correction: 30 µsec to 150 µsec
•
Linear Range with DTC: 1 mR/hr to 8 R/hr
Typical Checkpoints:
4
mR/hr
8
mR/hr
40 mR/hr
- calibration constant set point
80
mR/hr
400
mR/hr
800
mR/hr
4,000 mR/hr - dead time correction set point
8,000
mR/hr