7526A
Getting Started
20
TC Type
Range (
°
C)
Absolute Uncertainty, tcal ±5
°
C, ±(°C)
[1]
Output/Input
Minimum
Maximum
90 days
1 Year
T
-250
°
C
-200
°
C
-100
°
C
0
°
C
200
°
C
-200
°
C
-100
°
C
0
°
C
200
°
C
400
°
C
0.34
°
C
0.14
°
C
0.09
°
C
0.07
°
C
0.06
°
C
0.35
°
C
0.16
°
C
0.11
°
C
0.09
°
C
0.09
°
C
U
-200
°
C
0
°
C
200
°
C
0
°
C
200
°
C
600
°
C
0.15
°
C
0.08
°
C
0.07
°
C
0.16
°
C
0.10
°
C
0.10
°
C
XK
-200
°
C
-100
°
C
0
°
C
600
°
C
-100
°
C
0
°
C
600
°
C
800
°
C
0.10
°
C
0.07
°
C
0.06
°
C
0.07
°
C
0.11
°
C
0.09
°
C
0.08
°
C
0.09
°
C
BP
0
°
C
200
°
C
600
°
C
800
°
C
1600
°
C
2000
°
C
200
°
C
600
°
C
800
°
C
1600
°
C
2000
°
C
2500
°
C
600
°
C
0.17
°
C
0.14
°
C
0.15
°
C
0.22
°
C
0.26
°
C
0.38
°
C
800
°
C
0.18
°
C
0.16
°
C
0.17
°
C
0.23
°
C
0.28
°
C
0.40
°
C
[1]
Does not include thermocouple wire error.
Type B, E, J ,K, N, R, S and T are based on ITS-90
Type L and U are based on DIN 43710-1985
Type C is based on ASTM standard E 988-96
Type XK and BP are based on GOST R 8.585-2001
RTD and Thermistor Specification, Output
RTD Types
Range
°
C
Absolute Uncertainty, tcal
±
5
°
C
±
(
°
C)
[1]
Minimum
1 Year
90 Days
1 Year
Pt 385, 100
Ω
-200
°
C -800
°
C 0.04
°
C 0.05
°
C
Pt 3926, 100
Ω
-200
°
C 630
°
C 0.04
°
C 0.05
°
C
Pt 3916, 100
Ω
-200
°
C 630
°
C 0.04
°
C 0.05
°
C
Pt 385, 200
Ω
-200
°
C
400
°
C
400
°
C
630
°
C
0.35
°
C
0.42
°
C
0.40
°
C
0.50
°
C
Pt 385, 500
Ω
-200
°
C 630
°
C 0.15
°
C 0.17
°
C
Pt 385, 1000
Ω
-200
°
C 630
°
C 0.07
°
C 0.09
°
C
Ni 120, 120
Ω
-80
°
C 260
°
C 0.02
°
C 0.02
°
C
Cu 427, 10
Ω
[2]
-100
°
C 260
°
C 0.30
°
C 0.38
°
C
YSI 400
15
°
C 50
°
C 0.005
°
C 0.007
°
C
[1] 2-wire
output
[2]
Based on MINCO Application Aid No. 18
Summary of Contents for 7526A
Page 4: ...7526A Getting Started ii ...
Page 6: ...7526A Getting Started iv ...
Page 8: ...7526A Getting Started vi ...