Fluke Calibration 5322A-LOAD Instructions Manual Download Page 2

5322A-LOAD 

Instructions

 

 

Symbols 

The symbols used in these instructions and on 

the Product are shown in Table 

1

.  

Table 1. Symbols 

Symbol Description 

X

 

WARNING. HAZARDOUS VOLTAGE. Risk 
of electric shock. 

W

 

WARNING.RISK OF DANGER. Risk of 
electric shock. 

 

Consult user documentation. 

)

 

Certified by CSA Group to North American 
safety standards. 

P

 

Conforms to European Union directives. 

 

Conforms to relevant Australian Safety and 
EMC standards. 

 

AC (Alternating Current) 

.

 

Protective conductor terminal 

 

Earth 

 

Fuse 

 

Conforms to relevant South Korean EMC 
Standards. 

~

 

This product complies with the WEEE 
Directive marking requirements. The affixed 
label indicates that you must not discard this 
electrical/electronic product in domestic 
household waste. Product Category: With 
reference to the equipment types in the 
WEEE Directive Annex I, this product is 
classed as category 9 "Monitoring and 
Control Instrumentation" product. Do not 
dispose of this product as unsorted 
municipal waste. 

Preparing for Operation 

The Product (see Figure 2) comes with a power line 
adapter (to run the cooling fans), a ground cable, three 
shorting cables, and these instructions. 
The power line adapter comes with five mains plug 
adapters that are compatible with different power plug 
configurations.  
Prior to use: 
1.  Select the appropriate plug adapter that matches 

the local power mains outlet.  

2.  Attach the plug to the power line adapter. The 

power line adapter is now ready for use with the 
Product. 

3.  Once you place the Product on top of the 

Calibrator, connect the grounding cable between 
the Product’s rear-panel ground terminal and the 
rear-panel ground terminal of the Calibrator. See 
the grounding cable connection in Figure 3 where 
the rear of the Product is shown. 

XW

 Warning 

To prevent possible electrical shock, fire, 
or personal injury: 

 

Make sure that the ground terminal on 
the rear of the Product is connected to 
the ground (GND) terminal on the rear 
of the Calibrator. Any application with 
non-grounded housing is strictly 
prohibited.  

 

Use only the shrouded high-voltage 
banana plug jumpers that are supplied 
with this Product to connect resistor 
values. Connect the black terminal first 
and disconnect the black terminal last 
when you connect resistors. 

 

Summary of Contents for 5322A-LOAD

Page 1: ...this use the supplied high voltage jumper leads to short the resistor s red terminal to the resistor s corresponding black terminal The Product withstands a maximum voltage of 1 2 kV to 5 5 kV depend...

Page 2: ...gure 2 comes with a power line adapter to run the cooling fans a ground cable three shorting cables and these instructions The power line adapter comes with five mains plug adapters that are compatibl...

Page 3: ...nel switch to the I side of the switch to turn on the cooling fans A light on the front panel illuminates and indicates that power is on Turn off the power switch when the Product is no longer being u...

Page 4: ...ge levels can be safely applied for increasing longer times At specific levels sustained voltages can be applied indefinitely The other resistors 500k 1M 5M and 10M do not have a time limit for the ma...

Page 5: ...tors Versus Voltage Versus Current Current mA 500 V 1000 V 2000 V 5000 V 0 05 10M 0 1 5M 10M 0 2 5M 10M 0 3 5M 10M 0 4 5M 0 5 1M 10M 0 6 1M 5M 5M 10M 1 500k 1M 5M 1 5 500k 1M 5M 10M 2 250k 500k 1M 2 5...

Page 6: ...test voltage across each resistance between the respective input and COM terminals The resulting current through the resistor is measured and the value of the resistance is calculated Use a calibrato...

Page 7: ...0 20100 mA DCV ACV DCI ACI TRIG FREQ PERIOD TEMP ZERO INSTR SETUP MEAS SETUP RANGE F1 F2 F3 F4 F5 BACK MEMORY ANALYZE INPUT SENSE HI LO FRONT REAR 10A 1V 400 mA 4W HI LO 2W 4W V 300V FUSED 1000V CAT I...

Page 8: ...050 m 10 000 ft Storage 12 200 m 40 000 ft Dimensions 430 mm X 462 mm X 95 mm 16 9 in X 18 2 in X 3 7 in Weight net 3 kg 8 lb 4 5 oz Power Consumption 5 W maximum Safety class I according to EN 61010...