5322A-LOAD 

High Voltage Load Adapter 

Instructions 

 

Introduction 

The 5322A-LOAD High Voltage Load Adapter (the 
Product) loads a hipot tester to create leakage current 
during hipot tester calibration with the 5322A 
Multifunction Electrical Tester Calibrator (hereafter the 
Calibrator). As shown in Figure 1, the Product consists 
of nine power-rated resistors configured to provide 
nine resistance taps from 10 k

Ω

 to 10 M

Ω

. Each of the 

nine resistors in the Product can be put in parallel with 
any other resistor. To do this, use the supplied high-
voltage jumper leads to short the resistor's red terminal 
to the resistor’s corresponding black terminal.The 
Product withstands a maximum voltage of 1.2 kV to 
5.5 kV, depending upon the selected resistor.  

HI

LO

10k

25k

50k

100k

250k 500k

1M

5M

10M

Output

 

Ieu004.eps 

Figure 1. Load Schematic

 

 

 

XW

 

Warning 

To avoid possible electric shock or 
personal injury, use this Product only as 
specified in this Instruction Sheet or the 

5322A Operators Manual.

 

How to Contact Fluke Calibration 

To contact Fluke Calibration, call one of the following 
telephone numbers: 

  Technical Support USA: 1-877-355-3225 

  Calibration/Repair USA: 1-877-355-3225 

 Canada: 

1-800-36-FLUKE 

(1-800-363-5853) 

 Europe: 

+31-40-2675-200 

 Japan: 

+81-3-6714-3114 

 Singapore: 

+65-6799-5566 

  China:  +86-400-810-3435 

  Brazil:  +55-11-3759-7600 

  Anywhere in the world: +1-425-446-6110 

 

To see product information and download the latest 
manual supplements, visit Fluke Calibration’s website 
at 

www.flukecal.com

To register your product, visit 

http://flukecal.com/register-product

Summary of Contents for 5322A-LOAD

Page 1: ...this use the supplied high voltage jumper leads to short the resistor s red terminal to the resistor s corresponding black terminal The Product withstands a maximum voltage of 1 2 kV to 5 5 kV depend...

Page 2: ...gure 2 comes with a power line adapter to run the cooling fans a ground cable three shorting cables and these instructions The power line adapter comes with five mains plug adapters that are compatibl...

Page 3: ...nel switch to the I side of the switch to turn on the cooling fans A light on the front panel illuminates and indicates that power is on Turn off the power switch when the Product is no longer being u...

Page 4: ...ge levels can be safely applied for increasing longer times At specific levels sustained voltages can be applied indefinitely The other resistors 500k 1M 5M and 10M do not have a time limit for the ma...

Page 5: ...tors Versus Voltage Versus Current Current mA 500 V 1000 V 2000 V 5000 V 0 05 10M 0 1 5M 10M 0 2 5M 10M 0 3 5M 10M 0 4 5M 0 5 1M 10M 0 6 1M 5M 5M 10M 1 500k 1M 5M 1 5 500k 1M 5M 10M 2 250k 500k 1M 2 5...

Page 6: ...test voltage across each resistance between the respective input and COM terminals The resulting current through the resistor is measured and the value of the resistance is calculated Use a calibrato...

Page 7: ...0 20100 mA DCV ACV DCI ACI TRIG FREQ PERIOD TEMP ZERO INSTR SETUP MEAS SETUP RANGE F1 F2 F3 F4 F5 BACK MEMORY ANALYZE INPUT SENSE HI LO FRONT REAR 10A 1V 400 mA 4W HI LO 2W 4W V 300V FUSED 1000V CAT I...

Page 8: ...050 m 10 000 ft Storage 12 200 m 40 000 ft Dimensions 430 mm X 462 mm X 95 mm 16 9 in X 18 2 in X 3 7 in Weight net 3 kg 8 lb 4 5 oz Power Consumption 5 W maximum Safety class I according to EN 61010...

Reviews: