Maintenance & Servicing
Safety Valve
5
5-13
1
3
5
11
2
12
13
14
4
4
6
5
8
9
7
10
gjn36.eps
Figure 5-6. System Selector Valve
Table 5-6. System Selector Valve - Parts List
Item
Description
Part
Item
Description
Part
1
Set Screw
3927102
8
Valve Rotor
3921856
2 Knob
3927125
9 Screw
3908867
3
Valve Shaft
3921842
10
Bonded Seal
3921935
4 Pin
3906008
11 Connector 3868081
5 Bearing 3905259
12 Support
Bracket
3921492
6 Valve
Body
3921839
13 Screw
2687927
7 O-Ring
3905713
14 Spring
Plunger
3921874
Safety Valve
Disassembly
1.
Unscrew safety plug (3) from the vacuum body (7), taking care not to lose the
bonded seal (6).
2.
Carefully hold the larger “head” of the safety valve (1), and remove screw (5) and
valve stop (4).
3.
Withdraw safety valve (1) from safety plug (3), and inspect O ring (2) for cleanliness
or signs of damage—replace as necessary.
Re-Assembly
Reassembly is the reverse of the above disassembly instructions, taking care that all
sealing surfaces are clean and undamaged.