Upgrade the Connector Firmware

Note

For best results, make sure the battery is charged before you configure 
or load software on the Connector.

1.

On the PC, download the latest firmware from 

www.flukebiomedical.com

.

2.

Use the USB cable to attach the Connector to the PC.

3.

Turn on the Connector.

4.

On the PC, double-click the App icon.
a.

Click 

FIRMWARE UPGRADE

.

b.

Click 

Start Upgrade

.

c.

Browse to the latest firmware file. 

d.

Click 

Upgrade

.

Keep the device connected until the load completes.

5.

Turn off the Connector, then turn on the Connector.

Maintenance

XW

Warning

To prevent possible electrical shock, fire or personal injury, follow 
these guidelines:
• Repair the Product before use if the battery leaks.
• Be sure that the battery polarity is correct to prevent battery 

leakage.

• Use only Fluke approved power adapters to charge the battery.
• Batteries contain hazardous chemicals that can cause burns or 

explode. If exposure to chemicals occurs, clean with water and 
get medical aid.

• Do not disassemble the battery.
• Do not disassemble or crush battery cells and battery packs.
• Do not put battery cells and battery packs near heat or fire. Do 

not put in sunlight.

• Do not short the battery terminals together.
• Do not keep cells or batteries in a container where the terminals 

can be shorted.

• Do not operate the Product with covers removed or the case 

open. Hazardous voltage exposure is possible.

• Remove the input signals before you clean the Product.
• Use only specified replacement parts.
• Have an approved technician repair the Product.

Summary of Contents for WiFi Connector

Page 1: ...XWWarning To prevent possible electrical shock fire or personal injury follow these guidelines Read all safety information before you use the Product Carefully read all instructions Use the Product on...

Page 2: ...ty and EMC standards This product complies with the WEEE Directive marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household w...

Page 3: ...nnectivity B 802 15 1 or W WiFi USB port micro B for charging and upgrading firmware Power indicator Flashing green the Connector is on and not fully charged Solid green the Connector is fully charged...

Page 4: ...o the power adapter provided and turn on the Connector Install the Connector Application Before you use the Connector set up the WiFi Connector Utility the App on the PC 1 Download the App from www fl...

Page 5: ...and Device PIN Device PIN is the pairing code Follow the instructions in 802 15 1 Communications To configure WiFi communications click WiFi Settings Make a note of the WiFi SSID and the Password Fol...

Page 6: ...ks or Setup a new connection or network Manually connect to a wireless network 3 Make these settings Connect a Device to Ansur 1 Turn on the Connector Wait for the power indicator to show solid green...

Page 7: ...roduct before use if the battery leaks Be sure that the battery polarity is correct to prevent battery leakage Use only Fluke approved power adapters to charge the battery Batteries contain hazardous...

Page 8: ...5 Troubleshoot the Connector Problem Resolution Connector cannot communicate with the PC Make sure you used the Device PIN or WiFi password to pair the Connector to the PC Increase the baud rate of t...

Page 9: ...pacitive coupling for the treatment of material or inspection analysis purposes Class A equipment is suitable for use in nondomestic locations and or directly connected to a low voltage power supply n...

Page 10: ...THEORY This warranty covers only serialized products and their accessory items that bear a distinct serial number tag Recalibration of instruments is not covered under the warranty This warranty gives...

Reviews: