Vital Signs Simulator
Detailed Specifications
17
R-wave Detection
Waveform ............................................................ Triangular pulse
Rate ..................................................................... 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
Width ................................................................... 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
Width Accuracy ...................................................
±
(1 % of s 1 ms)
QRS Detection
Widths ................................................................. 8 ms to 20 ms in 2 ms steps and 20 ms to 200 ms in 10 ms steps
Width Accuracy ...................................................
±
(1 % of s 1 ms)
Rate ..................................................................... 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
R-Wave up slope ................................................. 0.875 amplitude, 0.4375 x width
R-Wave down slope ............................................ Full amplitude, 0.5 x width
S-Wave up slope ................................................. 0.125 amplitude, 0.0625 x width
Tall T-Wave Rejection
Waveform
QT Interval ...................................................... 350 ms
T-Wave width .................................................. 180 ms
T-Wave Shape ................................................ ½ sinewave
Amplitude ............................................................ 0 % to 150 % reference lead amplitude in 10 % steps
Rate ..................................................................... 80 BPM
Rate Accuracy
.......................................................
±
1 % of setting
Amplitude Accuracy
..............................................
±
(2 % of s 0.05 mV)