ProSim™ 8
Getting Started Manual
14
Amplitude Accuracy
.............................................
±
(2 % of s 0.05 mV)
ECG Rate
............................................................... 10 BPM to 360 BPM in 1 BPM steps
Rate Accuracy
.......................................................
±
1 % of setting
ECG Waveform Selection
..................................... Adult (80 ms) or pediatric (40 ms) QRS duration
ST-Segment Elevation
.......................................... Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps)
Additional steps: +0.05 mV and -0.05 mV
Power-On Default
.................................................. 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
Pacemaker Waveform
Pacer-Pulse
Amplitude
.......................................................... 0 mV (off),
±
2 mV,
±
4 mV,
±
6 mV,
±
8 mV,
±
10 mV,
±
12 mV,
±
14 mV,
±
16 mV,
±
18 mV,
±
20 mV,
±
50 mV,
±
100 mV,
±
200 mV,
±
500 mV, and
±
700 mV for lead II (reference lead)
Accuracy
Reference lead II .............................................
±
(5 % s 0.2 mV)
All other leads .................................................
±
(10 % s 0.4 mV)
Pacer-Pulse Width
................................................ 0.1 ms, 0.2 ms, 0.5 ms, 1.0 ms, and 2.0 ms
±
5 %
Paced Arrhythmias
............................................... Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-Ventricular sequential
Noncapture (one time)
Nonfunction
Power-On Default
.................................................. Amplitude 10 mV, width 1.0 ms, atrial waveform