FLUIDX VACUUM PUMP MANUAL - HIGH VACUUM ROTARY VANE PUMPS

FLUIDX VACUUM PUMP MANUAL - HIGH VACUUM ROTARY VANE PUMPS

2. INSTALLATION

2. INSTALLATION

2.3.1 Inlet port Connection

Connection Tip 

: To achieve specified pumping speed, make the length of pipeline between  

vacuum chamber and the pump as short as possible. Also the internal diameter of the pipeline is  

at least the same as inlet port; the narrow inlet line can reduce pumpingspeed.

You can protect the pump with applying proper accessories such as separators, filters and cooling

traps etc. between vacuum chamber and pump. However, the conductance value of the accessory

couldcause a reductioninpumpingspeed.

Connection

Connect the intake and exhaust lines with a clamp and centeringeach.
- Default : Connect to the NW25 flange supplied on the pump; use the clamp andcentering.

- Option 1 : The hose nipple can be provided upon request. Applicable to vacuum hose.  

(Ф 12 OD,  Ф 16 OD,  Ф 26 OD for NW25,  Ф 36 OD for NW40  available  depending  on hose size.)

- Option 2 : Screw nipple can be used. Applicable to screw valves etc.

(1/4

, 3/8

, 1/2

, 3/4

, 1

for NW25, 1&1/4” for NW40; available depending on connecting devices.)

Inlet filters can be installed for very dusty applications or pumping condensable vapor.
Valves can be used for isolating the pump from the vacuum system; to maintain vacuum when the

pump is switchedoff. Werecommendtoinstall the Angle Valve. (refer section7.2)

2.3.2 Outlet port Connection

WARNING

If the exhaust gases from the vacuum pump are harmful to human  

body, these must be safely lead away and subjected  to post-treatment  

as required

Connection Tip 

: Designing the exhaust line lower than the pump can prevent condensate  

from flowing back into the pump.
Oil vapors can be discharged from the exhaust port in low vacuum section (760 Torr~10 Torr)  

since the large amount of air is sucked into the vacuum pump. In order to reduce the emission of  

oil vapors, we recommend the installation of an additional exhaust filter (WSA oil mist trap : refer  

section 7.2) especially when the oil mist can not be lead out by separate hood or duct system.

Contact WSA for the detailedinquiry.

Connection

Exhaust connecting options are the same as Inlet Connection

*

Connection Tip

: Ensure that the internal diameter of the exhaust line is at least the same as outlet

port; the narrow exhaust line can accumulate the inner pressure in the pump and it can damage

the pump seals therefore cause oil leaks. The maximum pressure in the oil case must not exceed 1.5

bar (absolute).

CAUTION

If the temperature of the pump-body exceeds 70°C (158 °F),
you must fit suitable guards to prevent contact with hot surfaces

.

7

8

1. Unpack and Inspect

Remove all packing materials and protectivecovers.
Leave the inlet and outlet lids on before connection of the pump.  

(The pump is supplied filled withoil.)

Check the pump and accessories if they are the same type you have ordered.

Check the oil level contained through the sight glass.

Check the pump if it has  any sign of loose bolts, spilt oil, and broken/loose parts.

Do not use the pump if it is damaged; notify WSA with your order and serialnumber.

2. Location

The pump can be used free-standing on its baseplate.
To firmly install the pump in place, use bolts and fasten it to the ground or to a secure object  

through holes in the baseplate. It is recommended to use anti-vibration pads between base plate  

and fixing surface.
Locate the pump where you can check the oil level sight-glass and access the controls including  

oil filling-plug, oil drain-plug, andgas-ballast.
When installing the pump in a closed housing, make sure there is sufficient air circulation.
We use vacuum oil, ‘MR-200’ as our default vacuum oil and it is recommended to use thepump  

over 7 °C (45 °F)
WSA standard pump is not suited for installation in explosion hazard areas.  

If you need to apply the pump to such area, please contact us first.

2.3 Mechanical connection

Before connecting the WAS pump, remove the shipping seals from the inlet and exhaust  

connection flanges. (WSA pumps are sealed with white PE seals for domestic and ocean transport  

and metal blank clamped for air shipment to maintain oil inside)
WSA pumps are shipped with oil contained inside, so take extra care not to spill it  
Retain shipping seals in case you need to store the pump in the future.
Ensure that sealing surfaces are clean and scratch-free prior toassembly.

WARNING

Do not carry out any unauthorized conversions or  

modifications on the pump

CAUTION

The pump should be set up on a flat, horizontal surface.  

Terminal box should not to exposed to excessive moisture.

Summary of Contents for W2V10

Page 1: ...USER MANUAL HIGH VACUUM ROTARY VANE PUMPS W2V10 W2V20 W2V80 W2V40 W2V160...

Page 2: ...2 4 Electrical connection 3 Operation 10 3 1 Gas ballast control 3 2 Start up procedure 3 3 Decontamination of oil 3 3 1 Checking the oil level 3 3 2 Oil change 3 4 Switching O Shut down 4 Maintenance...

Page 3: ...tructions WARNING Indicatesproceduresthatmustbe strictly observed to preventpersonalinjury CAUTION Indicatesproceduresthatmustbe strictly observed to preventdamage topumporsystem Additional safety inf...

Page 4: ...Voltage Frequency Phase Full Load Power 0 4kW 0 5 HP Motor Speed 1 700 rpm Inlet and Outlet Options Standard NW25 Option I Hose Nipple 12 16 or 26 OD Option II PT Nipple 1 4 3 8 1 2 3 4 1 Option III...

Page 5: ...Request Voltage Frequency Phase Full Load Power 1 5kW 2 HP Motor Speed 1 700 rpm Inlet and Outlet Options Standard NW40 Change to NW25 available Option I Hose Nipple 36 OD for NW40 Option II PT Nippl...

Page 6: ...tionsare the same asInletConnection Connection Tip Ensure that the internal diameter of the exhaust line is at least the same as outlet port the narrow exhaust line can accumulate the inner pressure i...

Page 7: ...ifit stopsaccidentally Procedure Checkthatthe pump oil levelisbetweenthe Hand L marksthrough the oil levelsight glass if itis not refertosec 4 3 Neveroperate the pumpwiththe oil level belowL mark Tur...

Page 8: ...udo not youcan cause personal injury and property damage Allowthe pump tocool to a safe temperaturebeforeyoustartmaintenance work WARNING Disconnect the electrical connection beforedisassembling the p...

Page 9: ...oil level Daily 4 3 2Oil change Forproperoperationofthe pump itisessentialthatthe pump hasan adequate supplyofthe correctandcleanoil at all times Cleanvacuum oil isbasicallyodorlessandtransparent MR 2...

Page 10: ...s exposed to largeamounts of abrasivematerials a dust filtershould be fitted into the intakeline 4 5 Storage andDisposal Whena pump isputintooperationafteroverone yearof storage standardmaintenance sh...

Page 11: ...rotatesbut pump doesn trun Worn out coupling Replace coupling parts Worn out key between motor and pump Replace the key and setscrews Any foreign material blocks the pump Overhaul the pump The pump f...

Page 12: ...0 FLUIDX INC FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS 6 TECHNICAL DATA 6 TECHNICAL DATA 6 2 Development figure ofW2V10 6 1 Dimens...

Page 13: ...0 FLUIDX INC FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS 6 TECHNICAL DATA 6 TECHNICAL DATA 6 2 Development figure of W2V20 6 1 Dimen...

Page 14: ...0 FLUIDX INC FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS 6 TECHNICAL DATA 6 TECHNICAL DATA 6 2 Development figure of W2V40 6 1 Dimen...

Page 15: ...C FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS FLUIDX VACUUM PUMP MANUAL HIGH VACUUM ROTARY VANE PUMPS 6 TECHNICAL DATA 6 TECHNICAL DATA 6 2 Development figure of W2V80 6 1 Dimensions of W2...

Page 16: ...WPMIW5G00000 WASHER 5 GEAR SPRING 1 508 WPMIM5X20HN BOLT M5 X 20 1 509 WPRSP6000000 O RING P 6 SILICON 1 510 WPMI05102100 TAPER PIN ONE SCREW 7 X 25L M6 2 511 WPMIM6000000 NUT M6 2 512 WPMIM6X25HN BO...

Page 17: ...NDLE DIECAST 1 111 WPMIM6X15HN BOLT M6X15 2 112 WPMIM6X20HN BOLT M6X20 4 114 WPMC01140700 MOTOR PLATE GUIDE PIN 6 X 13L BRASS 2 115 WPMIM5X8HS00 BOLT M5X8 SET SCREW 2 116 WPMIM5X8HS00 BOLT M5X8 SET SC...

Page 18: ...2 X 6t ROUND 45x70x6 5 2 412 WPMS04122100 1ST STAGE BLADE SPRING W0 4 X 2 7 1 9 X 30L SUS 2 413 WPMI04132100 1ST STAGE BLADE SPRING GUIDE 1 8 3 X 26 2 3 x 26 2 414 WPME04140200 2ND STAGE BLADE 30 X 22...

Page 19: ...4 240 WPMIM4X8HB0 BOLT M4 X 8 ROUND HEAD Ni PLATED 4 Part No Code No Description Specification Qty 241 WPRE024101YW SIDE COVER WSA ABS YELLOW 2 301 WPMI03010400 1ST STAGE STATOR FC 25 1 302 WPMC030221...

Page 20: ...NGE NUT FOR B5 4 111 WPMI01111400 HOOK EYE BOLT M10X20 1 112 WPMIM8X30HN BOLT M8X30 4 113 WPMA01130800 FAN COVER 258 X 226 X 1 5T RM 1 114 WPMC01141400 MOTOR PLATE GUIDE PIN 8 X 20L BRASS 2 115 WPMIM6...

Page 21: ...6 X 41 X 7t ROUND39x88x7 5 2 415 WPMS04152100 2ND STAGE BLADE SPRING W0 4 X 2 7 1 9 X 30L SUS 2 416 WPMI04162100 2ND STAGE BLADE SPRING 1 8 3 X 26 2 3 x 26 2 501 WPMI05010800 2ND STAGE STATOR FC 25 88...

Page 22: ...6t Major only 34 415 1ST STAGE BLADE SPRING 2 W0 4 X 2 7 1 9 X 30L SUS Major only 35 416 1ST STAGE BLADE SPRING GUIDE 2 1 8 3 X 26 Major only 36 504 2ND VALVE SPRING 1 W0 6 X 7 3 8 5 X 17L Major only...

Page 23: ...STAGE BLADE 2 40 X 29 X 6t Major only 35 415 1ST STAGE BLADE SPRING 2 W0 4 X 2 7 1 9 X 30L SUS Major only 36 416 1ST STAGE BLADE SPRING GUIDE 2 1 8 3 X 26 Major only 37 504 2ND VALVE SPRING 2 W0 6 X...

Page 24: ...ccum hose fitting PT Nipple PT Nipple 1 4 3 8 1 2 3 4 1 forNW25 Screw Nipple Reducer Reducer VG50 NW25 VG150 NW25 NW25 NW16 NW40 NW25 NW50 NW40 Adjust In outlet size Angle Valve Air single bellows NW2...

Page 25: ...SMT 09 1001 Rev C1 CONTACT US PHONE FAX Phone 801 486 1015 Fax 801 486 1032 ADDRESS FLUIDX 139 W 2260 S Salt Lake City UT 84115 WEB www fluidxinc com info fluidxinc com...

Reviews: