FW_F126PEGTP_v1901_04_EN.docx
Page 40
Terminal connections
Type:
Removable plug-in terminal strip. Wire max. 1.5m m
2
and 2.5m m
2
Data protection
Type
EEPROM backup of all setting. Backup of running totals every minute.
Data retention at least 10 years.
Password
Configuration settings can be password protected.
Hazardous area
Intrinsically safe
Type XI
ATEX approval :
II 1 G Ex ia IIB/IIC T4 Ga
II 1 D Ex ia IIIC T
200
100°C Da
IECEx approval :
Ex ia IIB/IIC T4 Ga
Ex ia IIIC T
200
100°C Da
Explosion proof
Type XF
ATEX approval:
II 2 G Ex d IIB T5 Gb
II 2 D Ex t IIIB T100 °C Db
Weight appr. 15 kg.
Dimensions of enclosure: 350 x 250 x 200mm (13.7” x 9.9” x 7.9”) WxHxD.
Directives and
Standards
EMC
EN 61326-1; FCC 47 CFR part 15
LVD
EN/IEC 61010-1
ATEX / IECEx
EN/IEC 60079-0; EN/IEC 60079-11
RoHS
EN 50581
IP & NEMA
EN 60529; NEMA 250
Flowmeter
Type P
npn; npn-lp; reed; reed-lp; pnp; pnp-lp; namur; coil-hi; coil-lo; 8-1 DC; 12 DC; 24 DC
Frequency Minimum 0 Hz
–
maximum 7 kHz for total and flow rate.
Maximum frequency depends on signal type and internal low-pass filter.
E.g. Reed switch with low-pass filter: max. frequency 120 Hz.
K-Factor 0.000010 - 9999999 with variable decimal position.
Low-pass filter npn-lp; reed-lp; pnp-lp
Temperature
Type TP
2 or 3 wire PT100.
Standard range -100°C to +200°C (-148°F to 392°F) - accuracy 0.1 °C (0.18°F).
Extended range (ZV) -200°C to +800°C (-328°F to 1832°F) - accuracy 0.5 °C (0.9°F).
Update time four times a second.
The linearity is internally compensated.
Pressure
Type IA
(0)4-20mA; with signal calibration feature. Resolution: 14 bit.
Accuracy 0.05%; Low level cut-off programmable.
Span 0.000010 - 9999999 with variable decimal position.
Off-set 0.000 - 9999999 unit
Update time Four times a second.
Voltage drop 2.5 Volt @ 20mA.
Type IA: external power to sensor required; e.g. type PD.
Inputs