![Fluidwell F113-P-PD-OS Installation Manual Download Page 10](http://html1.mh-extra.com/html/fluidwell/f113-p-pd-os/f113-p-pd-os_installation-manual_2309008010.webp)
FW_F113POS_v1901_02_EN (addendum).docx
Page 10
4.4.3.
SENSOR SUPPLY
Option PD: Sensor supply: 3.2V, 8.2V, 12V or 24 V:
With this option, a real power supply for the sensor is available. The sensor can be powered with
8.2, 12 or 24 V DC (max. 50mA@24V). The voltage is selected by the three switches inside the
enclosure.
First, remove the terminal strip(s) after which the internal plastic cover can be removed. The
switches are located in the top left corner (option PD) as indicated:
Switch positions
SENSOR
A
VOLTAGE SELECTION
SWITCH
1
VOLTAGE
SWITCH
3
SWITCH
4
VOLTAGE
on
3.2 V DC
on
on
8.2 V DC
off
switch 3+4
off
on
12 V DC
on
off
12 V DC
off
off
23 V DC
Fig. 4: switch position voltage selection (option PD and PD-OS).
Function switch 1:
voltage selection sensor internal (3.2V) or external (switch 3+4).
Function switch 3+4:
the combination of these switches determine the voltage as indicated.
4.4.4.
TERMINAL CONNECTORS
The following paragraphs describe the terminal connections and configurations that are
specific for
type PD-OS (4 mechanical relay outputs)
.
For all other types of supply (PM, PF, PD, PX) in combination with types OA, OT and OR,
please consult the standard user manual
‘
FW_F113-P_vxxxx_xx_EN
’.
The following terminal connectors are available:
Fig. 5: Overview of terminal connectors - Standard configuration F113-P-OS and options
Terminal GND- 01- 02; power supply - only available with option PD:
O
PTION
SENSOR SUPPLY
T
erminal
backlight
GND
01
02
PD
24V AC
8.2, 12, 24V max 50mA
AC
AC
◊
PD
24V DC
8.2, 12, 24V max 50mA
L-
L+
◊
Note PD
do not use a AC autotransformer (Spartrafo) without a galvanic isolation.
◊
=option
on
off
4
3
1
INT EXT
1
INT E XT
on
off
PD-OS