Page 4
REFILL KIT, For use with Prometra
®
Programmable Infusion Systems
English
Instructions
Emptying the Drug Refill/Reservoir
7. Attach the syringe barrel to the stopcock. Place the syringe cap on the open en dof the syringe barrel.
Warning: Use only the calibrated syringe barrel supplied in the Refill Kit to collect infusate from the
drug reservoir.
Caution: Always make sure all connections are secure before addressing needle to skin. This will
prevent leaks.
8. Insert the needle through the center of the drug refill/reservoir access septum, perpendicular to
the skin and the pump. Advance the needle until the needle tip resides completely inside the refill
chamber.
Caution: Do not force the needle. Excessive force on the needle may damage the needle tip. Do not
rock the needle sideways as this may damage the septum or cause drug to leak into the pump pocket.
Emptying Setup
Emptying Setup in Pump