2001H Series Peristaltic Pump O&M Manual
7.
Clean inside of pump housing with damp rag or an
appropriate cleaning solution to remove any chemical
or tubing residue.
Reassembly
1. Reinstall the collet onto the pump shaft.
There is a slot in the collet that the flat drive tang on the
gearbox shaft must slide into. Orient the collet to allow the
drive tang to slide into the slot and push the collet
completely onto the gearbox shaft. When the collet
bottoms out it is in the correct position.
2. Mark an 11” section of tubing, which will be the portion, contained within the pump. Leave sufficient
excess on the suction and discharge sides of the pump for the desired connections. If you leave the
excess intake tubing in a coil near the pump it will make it easy to feed a new section of tubing through
the rollers when the section in the pump becomes worn.
Flomotion Systems, Inc. 2001H Series
Pg. 12
Mark an 11”
section of pump tubing.