5.9 Non-Contact AC Voltage Detector 

WARNING: It is possible for voltage to be present in a circuit even if the meter does not 

beep or flash the NCV LED lamp. Always verify meter operation on a known live AC current 
circuit and verify that the batteries are fresh before use. 

 

When the meter senses an AC Voltage or electromagnetic field > 100VAC, the following occurs: 

 

The meter produces audible tones (beeps) 

 

The LED lamp flashes ON and OFF 

 

The display shows 1, 2, 3, or 4 dashes 

The greater the electrical field strength, the faster the rate of the 
audible tones, the flashing of the LED lamp, and the number of dashes 
displayed. If the meter does not emit a tone or flash the LED, there is 
still the possibility that voltage is present; please use caution

1.

 

Turn the function switch to the 

 position to select Non-Contact 

Voltage Detect mode. 

2.

 

Note that EF is displayed when in this mode. If the EF does not 
display when the function switch is turned to the NCV position, 
check the batteries and do not use the meter until EF is displayed. 

3.

 

To test, place the meter near a source of electrical energy. Note 
that the tip of the meter offers the highest sensitivity. 

4.

 

Note the audible beeping, the flashing LED, and the displayed 
dashes when a source of electrical energy is detected. 

 

Fig 5-3 NON-CONTACT VOLTAGE DETECTOR 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for VT8-1000

Page 1: ...USER MANUAL Voltage Continuity and Current Testers VT8 600 and VT8 1000...

Page 2: ...TION 9 5 1 Meter Power ON OFF 9 5 2 Display Backlight 9 5 3 Worklight 9 5 4 Data Hold 9 5 5 Test Lead Considerations 10 5 6 Voltage Measurement Warnings and Cautions 10 5 7 AC DC Voltage Measurements...

Page 3: ...for identification only and are the property of their respective owners 1 2 Quality Assurance The Quality Management System under which these products are developed and manufactured has been certifie...

Page 4: ...ct Updates Product Registration and Customer Support 2 1 Features 6000 count 0 5999 digital display LCD display with high low intensity backlight True RMS AC measurements 0 8 DCV accuracy Data Hold fr...

Page 5: ...ment or meter housing Use the meter only as specified in this User Guide or accompanying Quick Manual to avoid compromising the protections provided by the meter Be sure to use the proper terminals sw...

Page 6: ...rms to EU directives ETL listed complies with USA and Canada requirements PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLT AGE CATEGORY I is equipment for con...

Page 7: ...l effect Current sensor 3 NCV alert LED lamp 4 Worklight button 5 Rotary function switch 6 REL Relative and ZERO button 7 MODE button 8 HOLD and LCD Backlight button 9 LCD Display 10 Test lead jacks 1...

Page 8: ...hort press to step through the functions shown in the table below Fig 4 3 MODE Button Function Table Switch Position Mode Button Function AC DC AC DC REL ZERO Short press to enter exit the Relative mo...

Page 9: ...e meter switches OFF after 15 minutes of inactivity The low battery symbol appears on the display when the battery voltage is insufficient 5 2 Display Backlight The LCD backlight has two intensity lev...

Page 10: ...moving the test leads remove the red before the black lead Disconnect probes and circuit under test after all measurements are completed 5 7 AC DC Voltage Measurements 1 Insert the black test lead ban...

Page 11: ...rom the meter 2 Set the function switch to the position 3 Use the MODE button to select AC or DC For DC press ZERO to remove any residual magnetism before testing a conductor 4 Test only one conductor...

Page 12: ...the audible tones the flashing of the LED lamp and the number of dashes displayed If the meter does not emit a tone or flash the LED there is still the possibility that voltage is present please use...

Page 13: ...l resistance 6 Read the resistance value in the display The display will indicate the proper decimal point and value If the reading is out of range the OL display icon will appear Fig 5 4 RESISTANCE M...

Page 14: ...cross the wire or circuit under test 5 If the resistance is 10 the beeper will sound continuously If the resistance is between 10 and 100 the beeper may or may not sound For an open circuit condition...

Page 15: ...s use the REL button Relative mode to remove test lead and probe capacitance 6 Read the capacitance value in the display 7 The display will indicate the proper decimal point and value Fig 5 6 CAPACITA...

Page 16: ...e meter is not to be used for 60 days or more remove the batteries and store separately Do not mix battery types or freshness levels please use batteries of the same type and of the same freshness lev...

Page 17: ...o null the display reading AC Voltage 6 000V 0 001V 1 2 5 digits 600V VT8 600 1000V VT8 1000 60 00V 0 01V 1 2 3 digits 600 0V 0 1V 1 5 2 digits 1000V VT8 1000 0 1V 1 5 3 digits True RMS voltage applic...

Page 18: ...Protection Capacitance 60 00nF 0 01nF 4 0 20 digits 600V DC VT8 600 1000V DC VT8 1000 600V AC 600 0nF 0 1nF 6 000 F 0 001 F 60 00 F 0 01 F 600 0 F 0 1 F 6 000mF 0 001mF 10 60 00mF 0 01mF For referenc...

Page 19: ...600V AC rms DC VT8 600 maximum applied to any terminal Auto Power OFF After 15 minutes Operating Temperature 14 122 F 10 50 C Storage Temperature 22 140 F 30 60 C Relative Humidity 95 maximum 41 86 F...

Page 20: ...ort flir com Service Repair Support Email Repair flir com Support Telephone number 1 855 499 3662 option 3 toll free 9 Three Year Limited Warranty This product is protected by FLIR s 3 Year Limited Wa...

Page 21: ...port flir com Technical Support Email TMSupport flir com Service and Repair Email Repair flir com Customer Support Telephone 1 855 499 3662 option 3 toll free Publication Identification No VT8 600_100...

Reviews: