Operation
5
5. If the thermocouple is not connected when the Type-K mode is selected,
the display will show dashes in place of a reading. If the measurement is
out of range, the display will show ‘
OL
’.
6. To find the optimum emissivity setting for a given surface, take an IR tem-
perature measurement and then take a Type-K measurement. Adjust the
emissivity until the IR measurement value equals the Type-K measure-
ment value. Now the emissivity is optimized. Emissivity can be set in the
menu system (under
Measurement
).
5.5
Visible Spectrum Camera
Figure 5.5
Visible Spectrum Digital Camera Image
1. Long press the power button to switch the camera ON.
2. Select the Visible Image mode in the menu system (under
Image Adjust-
ments/Image Mode
).
3. Point the camera toward the test area and scan as desired.
4. View the image on the display, see
Figure 5–5
. Pull the image capture
trigger to save an image. Refer to Section 5.6,
Capturing and Working with
Images
, (next) for more information.
5.6
Capturing and Working with Images
1. To capture a camera image, pull and release the trigger. Note that an im-
age cannot be saved if a USB cable is connected.
2. After a successful capture, a display confirmation will briefly appear show-
ing the filename for the image.
3. The image is now stored in the internal memory.