4226044 (Rev. C) © 2020 FLIR Systems, Inc.
33
FLIR Fido® X4 Operations Manual
Step 4:
The TD slot will automatically close and the message “Checking System Per-
formance” will be displayed along with a progress indicator.
Figure 5-7
: Checking System Performance screen
Caution:
The metal portions of the reference standard can be
HOT. Handle with care to avoid burns.
Step 5:
When the TD slot opens, remove the reference standard.
Step 6:
Close the reference standard by lightly pressing the thumb button and sliding
until the extension is fully retracted back into the sheath.
Step 7:
Store the closed reference standard in the kit.
Step 8:
When analysis is complete, the Fido X4 will automatically display the result:
“Verify Pass," "Verify Fail," or an error message will appear. If “Verify Pass”
appears (Figure 5-8), select
OK
.
Figure 5-8
: Verify Pass screen