FLIR Extech Instruments Oyster-10 Manual Download Page 7

Oyster-10, V1.9 11/09 

7

Maintenance and Troubleshooting 

Electrode Cleaning 

The type of electrode coating that occurs over time will determine the cleaning method. 
Soft coatings can be removed by vigorous stirring or by using a squirt bottle. Organic 
chemical coating (hard coatings) should be chemically removed. Only in extreme cases 
should the bulb be cleaned mechanically. 

Battery Replacement 

Install a new battery when the low battery indicator appears on the LCD. Remove the 
battery cover, install a new 9V battery, and replace the battery cover. 

Troubleshooting the Meter

 

Meters are the least likely cause of pH measurement problems. To rule out the meter 
check the following: 
1.  Check the 9V battery and replace it if necessary. 
2.  When using the AC adaptor for power, check that the line cord is firmly connected to 

the meter. 

3.  Replace the existing electrode connector with optional BNC shorting plug (shorts the 

ground ring to the center pin of the meter’s BNC jack) and check the following: 
a)  Set the measurement mode to pH. The LCD reading should be stable. 
b)  Adjust the calibration control for an LCD reading of 7.00. 
c)  Change to the mV measurement mode. The LCD should read 0 mV. 
d)  Note that an out of range reading is indicated by dashes ( -------). 

Summary of Contents for Extech Instruments Oyster-10

Page 1: ...adings can be conveniently stored for later recall The Oyster 10 is battery operated portable and includes a hinged display that can be adjusted to any viewing angle In addition the neck strap offers...

Page 2: ...ter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery icon appears on LCD Operating Temperature 41 F to 104 F 5 C to 40 C Storage Temperature 4 o F to 140 o F 20 o C to 60 o...

Page 3: ...e below The meter automatically recognizes all of the buffers listed in the table below and calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 O...

Page 4: ...ater 8 Place approx 1 ml of pH 4 solution or tap water in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required for mV measurements 2 Use the pH mV button to...

Page 5: ...ear 5 To clear the MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This...

Page 6: ...o a sample buffer or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible...

Page 7: ...d replace the battery cover Troubleshooting the Meter Meters are the least likely cause of pH measurement problems To rule out the meter check the following 1 Check the 9V battery and replace it if ne...

Page 8: ...fraction has to be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Lon...

Page 9: ...rance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification impr...

Reviews: