FLIR Extech Instruments Oyster-10 Manual Download Page 3

Oyster-10, V1.9 11/09 

3

pH Overview 

pH is a unit of measure (ranging from 0 to 14pH) indicating the degree of acidity or 
alkalinity of a solution. pH tests are the most commonly performed measurements in water 
analysis, using the hydrogen ion activity of a solution to determine acidity or alkalinity. 
Solutions with a pH less than 7 are considered acidic, solutions with a pH higher than 7 are 
known as bases, and solutions with a pH of exactly 7 are considered neutral. 
The pH scale is logarithmic, so, for example, if sample A is 1 pH less than Sample B, this 
means that Sample A is 10 times more acidic then Sample B. A difference of 1 pH 
represents a ten-fold difference in acidity. 

ORP/REDOX Overview 

ORP stands for 

O

xidation 

R

eduction 

P

otential and represents the oxidizing or reducing 

nature of a solution. The overall “reactive” tendency of a solution can be determined from 
ORP measurements. ORP testing is becoming more common in waste-water and water 
treatment measurement regimens. The unit of measurement for ORP is mV.

 

Calibration and Operation 

pH Calibration (NOTE: ORP does not require calibration)

 

Before accurate measurements can be obtained, it is necessary to calibrate the meter with 
the electrode. Typically, pH 7 is calibrated first and then one or more of the other buffers 
(shown in the left column in the table below). The meter automatically recognizes all of the 
buffers listed in the table below and calibrates itself to the buffer’s pH value. 

1.  Connect the pH electrode to the meter’s BNC connector input. 
2.  Open the top half of the meter to an appropriate viewing position. 
3.  Power the meter with the 

POWER/CAL

 button. 

4.  Place the electrode into a buffer solution (4, 7, or 10). 
5.  Press and hold the 

CAL

 key until the CAL display icon appears. 

6.  During calibration, CAL is shown on the display followed by SA

 

(saved). 

7.  When calibration is complete, the meter automatically displays END and returns to 

normal operation mode. Repeat steps 1-5 as necessary. Note that it is not necessary 
to turn the meter off and then back on again to calibration another reference. 

Important calibration note: 

Since pH electrodes can vary in performance, this meter 

includes a special feature that displays an electrode’s slope in %. For example, if two 
electrodes (one new and moderately used) are calibrated in the same pH 4 solution they’ll 
both display pH 4 on the meter. This is because the meter will accept one pH value for a 
range of measurement values as shown in the table below (in this case, any reading 
between 1.68pH and 6.50pH will be accepted as 4pH calibration). The meter then displays 
the slope % for the electrode under test. If the slope % is below 75% or above 125%, the 
electrode must be replaced. A slope of 100% is ideal. 

pH reading 

Range of measurement 

1.68 

0.00 to 1.68 

4.00 

1.68 to 6.50 

7.00 

6.50 to 7.50 

10.00 

7.50 to 12.45 

12.45 

12.45 to 14.00 

 

Summary of Contents for Extech Instruments Oyster-10

Page 1: ...adings can be conveniently stored for later recall The Oyster 10 is battery operated portable and includes a hinged display that can be adjusted to any viewing angle In addition the neck strap offers...

Page 2: ...ter Auto Power OFF After 10 minutes of inactivity Low Battery Indication Battery icon appears on LCD Operating Temperature 41 F to 104 F 5 C to 40 C Storage Temperature 4 o F to 140 o F 20 o C to 60 o...

Page 3: ...e below The meter automatically recognizes all of the buffers listed in the table below and calibrates itself to the buffer s pH value 1 Connect the pH electrode to the meter s BNC connector input 2 O...

Page 4: ...ater 8 Place approx 1 ml of pH 4 solution or tap water in the wetting cap before replacing it mV ORP Measurements 1 Note that calibration is not required for mV measurements 2 Use the pH mV button to...

Page 5: ...ear 5 To clear the MAX MIN memory cycle the meter power External Temperature Offset Adjustment The Offset Adjustment is used to fine tune the temperature reading of the external temperature probe This...

Page 6: ...o a sample buffer or rinse solution shake the electrode with a snap motion to remove residual drops of solution 6 Use part of the sample buffer to be tested next as a rinse solution whenever possible...

Page 7: ...d replace the battery cover Troubleshooting the Meter Meters are the least likely cause of pH measurement problems To rule out the meter check the following 1 Check the 9V battery and replace it if ne...

Page 8: ...fraction has to be less than 50 to assure reasonable readings Limit time of measurement Keep probe in the 7 0 buffer between readings Dehydrated membrane Read solution for dry bulb below Dry Bulb Lon...

Page 9: ...rance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification impr...

Reviews: