Thermographic measurement techniques
16.3
Reflected apparent temperature
This parameter is used to compensate for the radiation reflected in the object. If the
emissivity is low and the object temperature relatively far from that of the reflected it will
be important to set and compensate for the reflected apparent temperature correctly.
16.4
Distance
The distance is the distance between the object and the front lens of the camera. This
parameter is used to compensate for the following two facts:
• That radiation from the target is absorbed by the atmosphere between the object and
the camera.
• That radiation from the atmosphere itself is detected by the camera.
16.5
Relative humidity
The camera can also compensate for the fact that the transmittance is also dependent
on the relative humidity of the atmosphere. To do this set the relative humidity to the cor-
rect value. For short distances and normal humidity the relative humidity can normally be
left at a default value of 50%.
16.6
Other parameters
In addition, some cameras and analysis programs from FLIR Systems allow you to com-
pensate for the following parameters:
• Atmospheric temperature –
i.e.
the temperature of the atmosphere between the cam-
era and the target
• External optics temperature –
i.e.
the temperature of any external lenses or windows
used in front of the camera
• External optics transmittance –
i.e.
the transmission of any external lenses or windows
used in front of the camera
#T559891; r. AE/57789/57789; en-US
39
Summary of Contents for A310 ex Series
Page 1: ...User s manual FLIR A310 ex nbn Austria GmbH...
Page 2: ......
Page 3: ...User s manual FLIR A310 ex T559891 r AE 57789 57789 en US iii...
Page 4: ......
Page 6: ......
Page 24: ...Mechanical drawings 9 See next page T559891 r AE 57789 57789 en US 18...
Page 26: ...EC Type Examination Certificate 10 See next page T559891 r AE 57789 57789 en US 20...
Page 27: ......
Page 28: ......
Page 30: ......
Page 31: ......
Page 33: ......
Page 35: ......
Page 36: ......
Page 37: ...EC Declaration of conformity enclosure 14 See next page T559891 r AE 57789 57789 en US 31...
Page 39: ...Certiticate of conformity camera 15 See next page T559891 r AE 57789 57789 en US 33...
Page 40: ......
Page 49: ...This section contains the original manual from the manufacturer of the enclosure...
Page 99: ......