Junction Box Type KAK – Technical Manual, Rev. 1.00   August 2007

 

Page 3 of 4

I

NTRODUCTION AND TECHNICAL 

D

ATA

 

The Aluminium junction box is designed for the parallel connection of 4 load cells. 

 

Type 

No. of load cells 

Housing size 

Inputs 

Output 

KAK-4 

up to 4 

80 x 250 x 52 mm 

4x M16 

1x M16 

 

The junction box type KAK can be connected to the instrumentation with a shielded 6-wire signal cable. 
The corner correction is done with resistors and/or potentiometer. 

 

Housing material: 

Aluminium 

Protection class: 

IP66 

Cable connection: 

With clamping terminals 

Corner correction: 

By exchangeable resistors and/or potentiometer 

 

M

ECHANICAL 

I

NSTALLATION

 

Look for a mounting location which is more or less dry and protected from environmental stress. 
 

E

LECTRICAL 

C

ONNECTIONS

 

 

Figure 1: Dimensions in [mm] 

 

The connection sequence of the load cells should correspond to the corners of the scale, i.e 
Corner 1 = Load cell 1, Corner 2 = Load cell 2, etc. 
 

L

OAD 

C

ELL 

C

ABLE 

C

ONNECTION

 

First the cable gland (see in fig.1) must be loosened. Then you have to feed the load cell cable through the 
cable gland unless the shrink tube is fully disappeared in the box. The wires have to run below the printed 
circuit board and will be pulled back to the top at the upper end of the printed circuit board. Afterwards you can 
connect the cables to the clamping terminals as indicated below: 

 

Cable coulour 

Description 

Terminal designation 

yellow 

=  Cable shield 

Shield / Schirm 

red 

=  Signal –  (Output –) 

red / rot 

white 

=    () 

white /weiss 

black 

=  Excitation –  (Input –) 

black / schwarz 

(if applicable, brown)* 

=  Sense – 

black / schwarz 

green 

=  Exci (Input +) 

green / grün 

(if applicable, blue)* 

=  Sense + 

green / grün 

* if load cell is equiped with 6-wire conductor cable 

 

After all conductors have been clamped to the terminals, the cable glands must be tightened. Please verify 
that all cable glands are tight and the cable is fully stress relieved. 