Kabelanschlusskasten Typ KPK-4/L6 Technisches Handbuch, Rev. 1.03 April 2017
Junction Box Type KPK-4/L6 Technical Manual
Seite / Page 9/12
Introduction and technical Data
The Polyester junction box is designed for the parallel connection of 4 load cells with 6 wire cable
The junction box type KPK can be connected to the instrumentation with a shielded 6-wire signal cable.
The corner correction is done with resistors.
Housing material KPK : Polyester
Housing material KAK : Aluminium
Protect class
: IP 66
Cable connection
: With clamping terminals
Corner correction
: By exchangeable resistors
Overvoltage protection : Surge arrestors
Mechanical Installation
Look for a mounting location which should be dry and protected from environmental stress.
Electrical Connections KPK
– 4 / L6
Figure 1:
Dimensions in [mm]
The earth screw (see (2) in fig.1) must be connected to protective earth or you must establish an electrical
connection to the weigh modules for potential equalisation.
The connection sequence of the load cells should correspond to the corners of the scale, i.e.
Corner 1 = Load cell 1, Corner 2 = Load cell 2, etc.